8. GENErAL SPECIFICATIONS

8-1

IM 01R04B05-00E-E 3rd edition July 30, 2010 -00

all Rights Reserved. Copyright © 2005, Rota Yokogawa

8. GENErAL SPECIFICATIONS

Qnom is the water flow rate at about 1 bar pressure drop.
the flowmeter has an automatically low cut at 0.05% of
Qnom.

accuracy mass flow

:

Liquid :

± 0.1% of flow rate

± zero stability / flow rate *100%

(refer to table 2)

Gas (option /Ga) : ± 0.5% of flow rate

± zero stability / flow rate *100%

(refer to table 2)

accuracy volume flow

:

SQRt ( (mass flow error in %)² + (density error in %)²)

Please refer to sizing.

accuracy based on the frequency output includes the com-
bined effects of repeatability, linearity and hysteresis.

Repeatability for liquids: ± 0.05%

± (zero stability/2) / flow rate

*100%

Batch process :

above specified accuracy if the

batch process is >1 minute. For

shorter batch time (dt in s) the

accuracy decreases with the

square root of 60/dt

PErFOrMANCE SPECIFICATIONS

Model
- Remote detector RCCS30 to 33: 2 tubes, low flow design
- Remote detector RCCS34 to 39/XR : 2 tube design
- Remote field-mount converter RCCF31
- Integral type RCCt34 to 39/XR: 2 tube integral design

Fluid to be measured : Liquid, gas or slurries

Measurement items : Mass flow, density, temperature

and derived from these values:

concentration, volume flow and

net flow

Mass flow measurement

table 1: measuring range

Type

rCCS30 rCCS31 rCCS32 rCCS33

Qmax

t/h

0.1

0.3

0.6

1.5

Qnom

t/h

0.045

0.17

0.37

0.9

Type

rCCx34 rCCx36 rCCx38 rCCx39

rCCx39

/Ir

rCCx39

/Xr

Qmax

t/h

5

17

50

170

300

600

Qnom

t/h

2.7

10

32

100

250

500

table 2 : Zero Stability

Type rCCS30 rCCS31 rCCS32 rCCS33
kg/h

0.005

0.0085

0.019

0.045

Type rCCx34 rCCx36 rCCx38 rCCx39

rCCx39

/Ir

rCCx39

/Xr

kg/h

0.135

0.5

1.6

5

13

25

Pressure dependency

the stiffness of the RotaMaSS tubes is slightly line pressure

dependent. the static pressure effect of mass flow and den-

sity can be corrected by setting the static pressure manually via

menu.

table 3 : Static pressure effect on mass flow (not corrected)

Type

rCCS30 rCCS31 rCCS32

rCCS33

rCCx34

% of
rate
per bar

SS

----

----

----

----

0.00081

HC 0.00000

0.00012

0.00246

0.02094

0.00084

Type

rCCx36

rCCx38

rCCx39

rCCx39

/Ir

rCCx39

/Xr

% of
rate
per bar

SS

0.00346

0.00950

0.01058

0.02920

0.00740

HC

0.00336

0.00896

0.00808

0.01780

----

density measurement

adjustment with water and air at calibration temperature.

With option /K4 thermal stabilized.

For option /K6 see also “Special calibrations” on page 3.

Measuring range

: 0.3 kg/l to 5 kg/l (RCCx39,

RCCx39/IR and RCCx39/XR to 2 kg/l)

no density measurement for gas application

table 4: accuracy (at calibration conditions):

Type

Standard

Option /K4

Option /K6

RCCS30

0.008 g/cm³

------

------

RCCS31

0.004 g/cm³

0.001 g/cm³

------

RCCS32

0.004 g/cm³

0.001 g/cm³

0.0005 g/cm³

RCCS33

0.004 g/cm³

0.001 g/cm³

0.0005 g/cm³

RCCx34

0.003 g/cm³

0.001 g/cm³

0.0005 g/cm³

RCCx36

0.0022 g/cm³

0.001 g/cm³

0.0005 g/cm³

RCCx38

0.0015 g/cm³

0.001 g/cm³

0.0005 g/cm³

RCCx39

0.0015 g/cm³

0.001 g/cm³

0.0005 g/cm³

RCCx39/IR

0.0015 g/cm³

------

------

RCCx39/XR

0.0015 g/cm³

------

------

Repeatability:

- RCCS32-33, RCCx34-39/XR : ± 0.0005 g/cm³ (Std, /K4)

Static pressure effect:

Compensated if static pressure is set in the menu

Installation:

vertical, else correction term must be set in the software

Specification of high performance density measurement (option /K6):

ambient temp. range : -10°C to 50°C

Fluid temp. range : -50°C to 150°C

Minimum flow rate for specified accuracy:

- RCCx36 to RCCx39 : 700 kg/h

- RCCx34 : 140 kg/h

- RCCS33 : 90 kg/h

- RCCS32 : 37 kg/h

Maximum flow rate : Qnom

Repeatability : ± 0.0002 g/cm³

temperature measurement: ±0.5°C ±0.2% of reading

Density accuracy : ±0.0005 g/cm³ (liquids not aerated /

no gas in the liquid)

Process temperature influence :

0.000015 g/cm³ * abs(tfluid-20°C)

-1.0

-0.9

-0.8

-0.7

-0.6

-0.5

-0.4

-0.3

-0.2

-0.1

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1.0

0

50

100

150

Flow in % of Qnom

Er

ro

r i

n

%

F10.EPS