A-13 – Yokogawa RotaMASS 3-Series User Manual

Page 113

Advertising
background image

APPENdIX 1. LIST OF PArAMETErS FOr EACH BLOCK OF rOTAMASS

a-13

IM 01R04B05-00E-E 3rd edition July 30, 2010 -00

all Rights Reserved. Copyright © 2005, Rota Yokogawa

relative

Index

Index

Parameter Name

Factory default

Write

Mode

Explanation

105

2105

SLuG_DuRatIon

1 s

o/S

Indicates a slug duration time. When the
value of SLuG_aLaRM_SELECt is “not
apply”, users cannot change this value.
Range 0 to 120sec.

106

2106

aFtER_SLuG

Measured value

o/S

Selects the behaviour of the output mass
flow after slug alarm has happened. When
the parameter is selected
-“Measured Value”, the output mass flow is
the measured value.
-“Hold”, the output mass flow is the value
before the slug alarm happened.
When the value of SLuG_aLaRM_SE-
LECt is “not apply”, users cannot change
this value.

107

2107

DRIVE_GaIn_DaMPInG

1 s

o/S

Indicates a drive gain damping time to get
a constant drive gain level. When the value
of SLuG_aLaRM_SELECt is “not apply”,
users can not change this value. Range 0
to 200sec.

108

2108

EMPtY_PIPE_aLM_SEL

Inhibit

o/S

When the value is set to “Empty Pipe sel”
this parameter enables:
EMPtY_PIPE_CRIt (Crit < density),
aFtER_EMPtY_PIPE.

When the value is set to “Dual Seal Sel”,
this parameter enables:
EMPtY_PIPE_CRIt (Crit > density),
aFtER_EMPtY_PIPE.

When the value is “Inhibit” they do not per-
form.
If gas measurement (/Ga) is ordered, this
parameter is not available.

109

2109

EMPtY_PIPE_CRIt

0.0 kg/l

o/S

Indicates the empty pipe criteria when
EMPtY_PIPE_aLM_SELECt is set to
“Empty Pipe Sel” (Crit < actual density).

Indicates the dual seal criteria when
EMPtY_PIPE_aLM_SELECt is set to “Dual
Seal Sel” (Crit > actual density).

110

2110

aFtER_EMPtY_PIPE

Mass flow = zero

o/S

Selects the behaviour of the output mass
flow after empty pipe alarm or dual seal
alarm has happened. When the parameter
is selected
-“Mass flow=zero”, the output mass flow is
zero.
-“Measured Value”, the output mass flow is
the measured value.
-“Hold”, the output mass flow is the value
before the empty pipe alarm happened.
When the value of EMPtY_PIPE_aLM_SE-
LECt is “not apply”, users cannot change
this value.

111

2111

CoRRoSIon_aLM_SEL

not apply

o/S

this parameter enables the functions of
CoRRoSIon_CRIt and CoRRoSIon_
DaMP.
When this value is “not apply” they do not
perform.
When this value is “apply” they do perform.
If gas measurement is ordered, this parameter
is not available. the value is “not apply”.

112

2112

CoRRoSIon_CRIt

1.5 kg/l

o/S

Indicate the corrosion criteria value. When
the value of CoRRoSIon_aLM_SEL is
“not apply”, users cannot change this value.
Range 0 to 5kg/l. unit depends on
DEnSItY_VaLuE_RanGE units index.

Advertising