APPENdIX 1. LIST OF PArAMETErS FOr EACH BLOCK OF rOTAMASS

a-11

IM 01R04B05-00E-E 3rd edition July 30, 2010 -00

all Rights Reserved. Copyright © 2005, Rota Yokogawa

relative

Index

Index

Parameter Name

Factory default

Write

Mode

Explanation

79

2079

REFEREnCE_DEnSItY

1.0 kg/l

o/S

this parameter indicates a reference density
value.
When the value of DEnSItY_FIX_VaL is
“Inhibit” users can not change this value and
this function does not perform. When the
value is “Enabled” users can change this
value and this parameter is used as reference
density for calculation of volume flow.

80

2080

DEnSItY_oFFSEt

0 g/l

o/S

this parameter indicates a density offset
value.

81

2081

tEMP_FIX_VaL_SELECt

Inhibit

o/S

this parameter enables the function of
tEMP_FIXED_VaLuE.
When the value is “Inhibit”, the function of
tEMP_FIXED_VaLuE does not perform.
When the value is “Enable”, the function of
tEMP_FIXED_VaLuE performs.

82

2082

tEMP_FIXED_VaLuE

30°C

o/S

this parameter indicates a fix temperature
value.
When the value of tEMP_FIX_VaL is
“Inhibit” users can not change this value and
this function does not perform. When the
value of tEMP_FIX_VaL is “Enable” users
can change this value and this parameter is
set to the temperature value.

83

2083

tEMP_GaIn

1.00

o/S

this parameter defines the value of tem-
perature gain. the function allows correction
of the temperature measurement for better
accuracy in mass flow and density for very
high and low temperatures.

84

2084

SEnSoR_MoDEL

*)

this parameter defines the sensor model.
When a user changes the value the follow-
ing parameters return to default automati-
cally:
SEnSoR_RanGE,
MaSS_FLoW_VaLuE_RanGE,
VoLuME_FLoW_VaLuE_RanGE,
nEt_FLoW_VaLuE_RanGE,
SK20, SKt, RV, QnoM,
KD, FL20, FtC1, FtCK,
SKP, FPC, SKtK, SKPt,
FPtC, FQC1, FQC2,
autoZERo_RanGE,
FLuCtuatIon_RanGE

85

2085

SK20

*)

o/S

Defines the sensor constant at 20°C.
the valid range and the default value de-
pend on the selected sensor model.

86

2086

SKt

*)

o/S

Defines a linear temperature coefficient of
the sensor constant SK20.
the valid range and the default value de-
pend on the selected sensor model.

87

2087

RV

*)

o/S

Defines a density correction factor.
the valid range and the default value de-
pend on the selected sensor model.

88

2088

QnoM

*)

o/S

Defines the water flow rate of about
1 bar pressure drop.
the valid range and the default value de-
pend on the selected sensor model.

89

2089

KD

*)

o/S

Defines a density calibration constant.
the valid range and the default value de-
pend on the selected sensor model.

90

2090

FL20

*)

o/S

Defines a response frequency of the tubes
filled with air at 20°C.
the valid range and the default value de-
pend on the selected sensor model.

*) Depends on detector size. For RCCF31 not combined with detector, data for RCCS36 are stored.