A-20 – Yokogawa RotaMASS 3-Series User Manual
Page 120

APPENdIX 1. LIST OF PArAMETErS FOr EACH BLOCK OF rOTAMASS
a-20
IM 01R04B05-00E-E 3rd edition July 30, 2010 -00
all Rights Reserved. Copyright © 2005, Rota Yokogawa
Table A2.1 INTEG_TyPE
name
Integration Method
Integration Range
Reset trigger (Reset if one of the
following conditions is established)
trip output
uP_auto(1) Counting up
Starting from "0"
-InF< total <totaL_SP
0< atotal <+InF
0< Rtotal <+InF
-InF< acctotal <+InF
• out reaches totaL_SP.
• RESEt_In = 1
• oP_CMD_Int = 1
s
uP_DEM(2)
Counting up
Starting from "0"
-InF< total <+InF
0< atotal <+InF
0< Rtotal <+InF
-InF< acctotal <+InF
• RESEt_In = 1
• oP_CMD_Int = 1
s
Dn_auto(3)
Counting down
Starting from
totaL_SP
0< total <+InF
0< atotal <+InF
0< Rtotal <+InF
-InF< acctotal <+InF
• out reaches "0."
• RESEt_In = 1
• oP_CMD_Int = 1
s
Dn_DEM(4)
Counting down
Starting from
totaL_SP
-InF< total <+InF
0< atotal <+InF
0< Rtotal <+InF
-InF< acctotal <+InF
• RESEt_In = 1
• oP_CMD_Int = 1
s
PERIoDIC(5) Counting up
Starting from "0"
-InF< total <+InF
0< atotal <+InF
0< Rtotal <+InF
-InF< acctotal <+InF
• at the period specified by
CLoCK_PER
• oP_CMD_Int = 1
3
DEManD(6)
Counting up
Starting from "0"
-InF< total <+InF
0< atotal <+InF
0< Rtotal <+InF
-InF< acctotal <+InF
• RESEt_In = 1
• oP_CMD_Int = 1
3
PER&DEM(7)
Legend s: trip output is made. 3: no trip output is made.
Counting up
Starting from "0"
-InF< total <+InF
0< atotal <+InF
0< Rtotal <+InF
-InF< acctotal <+InF
• at the period specif. by CLoCK_PER
• RESEt_In = 1
• oP_CMD_Int = 1
3
ta0203.EPS