A-10 – Yokogawa RotaMASS 3-Series User Manual

Page 110

Advertising
background image

APPENdIX 1. LIST OF PArAMETErS FOr EACH BLOCK OF rOTAMASS

a-10

IM 01R04B05-00E-E 3rd edition July 30, 2010 -00

all Rights Reserved. Copyright © 2005, Rota Yokogawa

relative

Index

Index

Parameter Name

Factory default

Write

Mode

Explanation

54

2054

DISP_DECIMaL_MaSS_FLoW

xxxx.XX

-

the decimal point position of the mass flow
value on the LCD display.

55

2055

DISP_DECIMaL_VoLuME_FLoW

xxxx.XX

-

the decimal point position of the volume
flow value on the LCD display.

56

2056

DISP_DECIMaL_DEnSItY

xx.XXXX

-

the decimal point position of the density
value on the LCD display.

57

2057

DISP_DECIMaL_tEMPERatuRE

xxxxx.X

-

the decimal point position of the temperature
value on the LCD display.

58

2058

DISP_DECIMaL_ConCEntR_MEaS xxxx.XX

-

the decimal point position of the concentration
measurement value on the LCD display.

59

2059

DISP_DECIMaL_nEt_FLoW

xxxx.XX

-

the decimal point position of the net flow
value on the LCD display.

60

2060

DISP_DECIMaL_It1

xxxxx.XX

-

the decimal point position of the out value
of It1 block on the LCD display.

61

2061

DISP_DECIMaL_It2

xxxxx.XX

-

the decimal point position of the out value
of It2 block on the LCD display.

62

2062

DISP_It1_unItS_InDEX

none

-

the unit index of the out value of It1 block
on the LCD display.

63

2063

DISP_It2_unItS_InDEX

none

-

the unit index of the out value of It2 block
on the LCD display.

64

2064

DISP_ContRaSt

4

-

the contrast of the LCD display.

65

2065

DISP_PERIoD

1 s

-

the update cycle of the LCD display.

66

2066

DISP_LanGuaGE

English

-

the language on the LCD display.

67

2067

FLoW_DIRECtIon

Forward

o/S

Direction of flow

68

2068

BI_DIRECtIon

Bi-direction

o/S

Selects the Bi-direction mode (bi-direction/
uni-direction).

69

2069

auto_ZERo_tIME

3 min

o/S

Defines the execution time of the autozero
function.

70

2070

auto_ZERo_EXE

not Execute

o/S

a user can execute the autozero performing
by this parameter.
after autozero execution is completed, this
value returns to “not Execute” automatically.
the execution time is defined by auto_
ZERo_tIME.

71

2071

auto_ZERo_VaLuE

Set after adjust-
ment

-

Indicates the result value of the autozero
execution.

72

2072

auto_ZERo_FLuCtuatIon

Set after adjust-
ment

-

Indicates the factory autozero range.

73

2073

aZ_InIt_MaSS_FLoW

Set after adjust-
ment

-

Indicates the last autozero value for mass
flow.

74

2074

aZ_InIt_DEnSItY

Set after adjust-
ment

-

Indicates the density at the last autozero for
mass flow.

75

2075

aZ_InIt_tEMP

Set after adjust-
ment

-

Indicates the temperature at the last
autozero for mass flow.

76

2076

MaSS_FLoW_FIX_VaL_SEL

Inhibit

o/S

this parameter enables the function of
MaSS_FLoW_FIXED_VaLuE.
When the value is “Inhibit”, the function of
MaSS_FLoW_FIXED_VaLuE does not
perform.
When the value is “Enable”, the function of
MaSS_FLoW_FIXED_VaLuE performs.

77

2077

MaSS_FLoW_FIXED_VaLuE

0.0 t/h

o/S

this parameter indicates a fix mass flow
value. the set value is added to the
autozero value.

78

2078

DEnSItY_FIX_VaL_SEL

Inhibit

o/S

this parameter enables the function of
REFEREnCE_DEnSItY.
When the value is “Inhibit” the function does
not perform. When the value is “Enabled”
the function performs.
If gas measurement (/Ga) is ordered, this
parameter is “Enabled”.