A-14 – Yokogawa RotaMASS 3-Series User Manual

Page 114

Advertising
background image

APPENdIX 1. LIST OF PArAMETErS FOr EACH BLOCK OF rOTAMASS

a-14

IM 01R04B05-00E-E 3rd edition July 30, 2010 -00

all Rights Reserved. Copyright © 2005, Rota Yokogawa

relative

Index

Index

Parameter Name

Factory default

Write

Mode

Explanation

113

2113

CoRRoSIon_DaMP

10 h

o/S

Indicates the corrosion damping time.
When the value of CoRRoSIon_aLM_SEL
is “not apply”, users cannot change this
value. Range to 0 to 10 hours.

114

2114

FLuID_MaX_tEMP

0.0

-

this parameter indicates the maximum
reached fluid temperature since the last
reset.
user cannot reset this value, only in service.

115

2115

SELF_tESt

not Execute

o/S

a user can execute the SELF_tESt function
by this parameter. after executing the ERR_
StatuS, aLM_StatuS and WaRnG_
StatuS parameter will be updated.

116

2116

InItIaLIZE_EEPRoM

not Execute

o/S

a user can execute the initialization of EEP-
RoM by this parameter.
after a user enters a password, parameters
return to default values.

117

2117

ERR_StatuS

-

-

this parameter indicates the actual error
situation in RCCt3/FB.

118

2118

aLM_StatuS

-

-

this parameter indicates the actual alarm
situation in RCCt3/FB.

119

2119

WaRnG_StatuS

-

-

this parameter indicates the actual warning
situation in RCCt3/FB.

120

2120

HISt_oRD

-

-

this parameter indicates the last 10 errors,
alarms, warning events/status in their order
of occurrences.
the oldest will be removed after an 11th
occur (first in-first out function).
this parameter can be cleared by CLEaR_
HISt Parameter.

121

2121

HISt_aBS_ERR

-

-

this parameter indicates all error events/
status absolute after occurrences.
this parameter can be cleared by CLEaR_
HISt.

122

2122

HISt_aBS_aLM

-

-

this parameter indicates all alarm events/
status absolute after occurrences.
this parameter can be cleared by CLEaR_
HISt parameter.

123

2123

HISt_aBS_WaRnG

-

-

this parameter indicates all warning events/
status absolute after occurrences.
this parameter can be cleared by CLEaR_
HISt parameter.

124

2124

CLEaR_HISt

not Execute

o/S

this parameter clear entries in the HISt_oRD,
HISt_aBS_ERR, HISt_aBS_aLM, and
HISt_aBS_WaRnG parameter.

125

2125

aLaRM_PERFoRM

-

-

this parameter clears alarm/Warning
information about each function block,
temporarily.
If a user writes 0 to a/some with this param-
eter, the corresponding alarm and warning
will be cleared.

126

2126

aLaRM_SuM

0

-

the parameter shows current alarm.

127

2127

REFEREnCE_tEMPERatuRE

25°C

o/S

Indicates the temperature where the
reference density of both components of the
mixture has been determined.

128

2128

REFEREnCE_DEnSItY_CaRRIER 0.997039 kg/l

o/S

Indicates the density of the carrier liquid de-
termined at the reference temperature.

129

2129

tEMP_CoEFF_a_CaRRIER

-0.261 E-3/K

o/S

Indicates the linear temperature
coefficient of the density of the carrier liquid.

130

2130

tEMP_CoEFF_B_CaRRIER

-0.36 E-5/K²

o/S

Indicates the quadratic temperature coefficient
of the density of the carrier liquid.

131

2131

REFEREnCE_DEnSItY_PRoDuCt 0.0 kg/l

o/S

Indicates the density of the product
determined at the reference temperature.

132

2132

tEMP_CoEFF_a_PRoDuCt

0.0 E-3/K

o/S

Indicates the linear temperature
coefficient of the density of the product.

133

2133

tEMP_CoEFF_B_PRoDuCt

0.0 E-5/K

2

o/S

Indicates the quadratic temperature coefficient
of the density of the product.