A-76, Table a8.2 transducer status, Figure a8.2 transducer status – Yokogawa RotaMASS 3-Series User Manual

Page 176

Advertising
background image

APPENdIX 8. dEVICEVIEWEr WINdOW EXECUTEd FrOM FIELdMATE ANd PrM

a-76

IM 01R04B05-00E-E 3rd edition July 30, 2010 -00

all Rights Reserved. Copyright © 2005, Rota Yokogawa

Table A8.2 Transducer Status

Frequency Failure(E-01)

Error

Driving frequency out of range. (E-01). (f < Frequency low limit, or f > Frequency high limit)
[Remedy]: Check cable (remote) and detector.

Signal Failure(E-02)

Error

Phase difference out of range. (E-02).
(deltaPHI < Phase Difference low limit, or deltaPHI > Phase Difference high limit)
[Remedy]: Check cable (remote) and detector.

Sensor 1 Signal Failure(E-06)

Error

Sensor 1 signal too small. (E-06). (S1 < 7% of Drive amplitude for 3 min)
[Remedy]: Check cable (remote) and detector.

Sensor 2 Signal Failure(E-07)

Error

Sensor 2 signal too small. (E-07). (S2 < 7% of Drive amplitude for 3 min)
[Remedy]: Check cable (remote) and detector.

temp Sensor Failure(E-08)

Error

temperature sensor failed. (E-08). (t<-210degC, or t>450degC)
[Remedy]: Check cable (remote) and detector. Check temperature.

SoftDL Incomplete(E-09)

Error

Software download is incomplete. (E-09)
[Remedy]: Check the cables, power and RB Softdown Error(RB.SoFtDWn_ERRoR).

SoftDL Failure(E-10)

Error

Software download failed. (E-10)
[Remedy]: Check the download file and RB Softdown Error(RB.SoFtDWn_ERRoR).

abnormal Boot Process

Error

abnormal boot processing was detected at the time of starting.
[Remedy]: Check the cables, power and RB Softdown Error(RB.SoFtDWn_ERRoR).

Slug Detection(a-14)

alarm

Slug criterion is reached. (a-14)
[Remedy]: Reduce gas bubbles in medium.

Empty Pipe Detection(a-15)

alarm

Empty Pipe criterion is reached. (a-15)
[Remedy]: Fill pipe.

Corrosion Detection(a-16)

alarm

Corrosion criterion is reached. (a-16)
[Remedy]: Change detector.

Density Lower 0.3kg/l(W-01)

Warning

Density lower than 0.3kg/l. (W-01)
[Remedy]: Fill detector with fluid.

Fixed Dens Selected(W-02)

Warning

Reference density is enabled. (W-02)
[Remedy]: Set tB Density Fixed Value Selection (tB.DEnSItY_FIX_VaL_SEL) to inhibit.

Fixed temp Selected(W-03)

Warning

temperature fixed value is enabled. (W-03)
[Remedy]: Set tB temperature Fixed Value Selection (tB.tEMP_FIX_VaL_SELECt) to inhibit.

Fixed Mass Flow Selected(W-04)

Warning

Mass flow fixed value is enabled. (W-04)
[Remedy]: Set tB Mass Flow Fixed Value Selection (tB.MaSS_FLoW_FIX_VaL_SEL) to inhibit.

autozero Value out of Range(W-6)

Warning

autozero value is out of the internal detector depending range. (W-06)
[Remedy]: Stop flow during autozero. Check detector installation.

autozero Fluct out of Range(W-7)

Warning

autozero fluctuation is out of the internal detector depending range. (W-07)
[Remedy]: Stop flow during autozero. Check electrical installation, vibration and density.

PD/Freq Simulation active(W-08)

Warning

Fixed frequency or fixed phase difference is enabled. (W-08)
[Remedy]: Set tB Sensor Simulation (tB.SEnSoR_SIMuLatIon) to inhibit.

Figure A8.2 Transducer Status