Electrical data of the DO202S.

- Supply circuit (terminals + and -):

Maximum input voltage Vmax = 31.5 V. Maximum input current Imax = 100 mA.

Maximum input power Pi = 1.2 W.

Effective internal capacitance Ci = 22 nF. Effective internal inductance Li = 35 PH.

- Sensor input circuit (terminals 11 through 18):

Maximum output voltage V

t

= 14.4 V. Maximum output current I

t

= 10 mA.

Maximum allowed external capacitance C

a

= 643 nF.

Maximum allowed external inductance L

a

= 340 mH.

・

If Hand Held Terminal (HHT) is not connected to the power supply lines of the DO202S

(see figure 1):

Any FM Approved barrier or power supply may be used that meets the following requirements.

V

oc

or V

t

d

31.5 V; I

sc

or I

t

d

100 mA; C

a

t

22nF + C

cable

; L

a

t

35PH + L

cable

If HHT is connected to the power supply lines of the DO202S (see figure 2):

The Hand Held Terminal must be FM Approved. Refer to the manufacturers control drawing of the

HHT and the barrier/power supply to determine the cable parameters.

(V

oc

or V

t

) + V

HHT

d

31.5 V; (I

sc

or I

t

) + I

HHT

d

100 mA;

C

a

t

22nF + C

cable

+ C

HHT

; L

a

t

35PH + L

cable

+ L

HHT

When installing this equipment, follow the manufacturer ’s installation drawing.

Installation should be in accordance with ANSI/ISA RP 12.06.01 “Installation of Intrinsically Safe

Systems for Hazardous (Classified) Locations” and the National Electrical Code (ANSI/NFPA 70).

Control equipment connected to the barrier/power supply must not use or generate more than 250

Vrms or Vdc.

・

Resistance between Intrinsically Safe Ground and earth ground must be less than1.0 Ohm.

・

In case of using cable glands in Outdoor location, they shall be UV rated or made of metal.

WARNING

- Substitution of components may impair Intrinsic Safety

- To prevent ignition of flammable or combustible atmospheres, disconnect power before servicing or

read, understand and adhere to the manufacturer ’s’live maintenance procedures.

Sensor
terminals 11-18
Max. cablelength: 60 mtr.
Cable dia. : 3…12 mm.

Sensor
terminals 11-18
Max. cablelength: 60 mtr.
Cable dia.: 3…12 mm.

Ùnclassified Location

FM Approved

Power Supply

(HART compatible)

Output

Supply

+

_

G

Classified Location

Unclassified Location

+

_

G

Functional
earth

Functional
earth

Intrinsically safe design
FM Class I, Div.1, Group ABCD,

T4 for ambient temp. < 55°C

T6 for ambient temp. < 40°C

DO202S transmitter

+

-

Load

Resistance

Classified Location

Functional
earth

+

_

24 volts DC Nominal

Supply Voltage.

FM Approved safety barrier or
power supply
with Rint = 300

:

(HART compatible)

For electrical data:
see text below.

For electrical data:
see text below.

Intrinsically safe design
FM Class I, Div.1, Group ABCD,

T4 for ambient temp. < 55°C

T6 for ambient temp. < 40°C

DO202S transmitter

Figure 1

Figure 2

General Specifications 2-8

2-6. Control Drawing of DO202S mA HART® Specification (FM Intrinsically safe design)