Appendix 1-1

A-1-1. User setting table

FUNCTION

SETTING DEFAULTS

USER SETTINGS

Parameter specific functions
01 *S.TYPE 1

Polarographic

*V.POL 0.675 V
*SENS 7.5

nA/ppm

04 *SAL.TY 0

Off

*[Cl-] 0

g/kg

Temperature measuring functions
10 *T.SENS 1

22k

NTC

11 *T.UNIT 0

°C

12 *T.ADJ

None

13 *T.MAN 0

Off

Calibration parameter functions
20 *∆T.SEC 60

sec.

*∆Z.PPM 0.1

ppm

*∆Z.PPB 1

ppb

*∆Z.SAT 1.0

%

*∆S.PPM 0.1

ppm

21 *∆Z.CAL 0

Off

*∆Z.LIM 199.9 nA

(polar)

22 *ZERO

0

μA (nA)

*SENS 7,5

nA/ppm

mA outputs
31 *OUTP.F 0

Linear

ppm

32 *BURN

0

No

Burn

35

*TABLE

0 to 20

ppm

User Interface
50 *RET

1

on

51 *P.MAN 0

off

*P.% 101.3 kPa
*P.CAL 101.3 kPa
52 *PASS

0.0.0

all

off

53 *Err.01

1

hard

fail

*Err.02 1

hard

fail

*Err.07 1

hard

fail

*Err.08 1

hard

fail

*Err.09 1

hard

fail

*Err.16 0

soft

fail

55 *CALL.M 0

Off

*CALL.M 250

days

56 *UNIT

0

ppm

Communication
60 *COMM. 1.0

on/write

enable

*ADDR. 00
61 *HOUR
62 *ERASE
General
70 *LOAD
Test and setup mode
80 *TEST

mA

mA

mA

APPENDIX 1. USER SETTINGS