IM 12J05C01-01E

2-12 Appendix

3/3

QIS 12J05C01-61E

i. Press

the

[ ] key. The instrument returns to the measurement mode.

Set the display to nA current indication by following the steps below.

a. Press the [MODE] key. The message display will show “CALIB.”

b. Press the [NO] key once. The message display will show “DISP.”

c. Press the [YES] key once. The message display will show temperature indication.

d. Press the [NO] key five times. The message display will show nA current indication.

Change the current of DC generator 2 as shown in Table 3. The corresponding indication must

be within the range.

Table 3

Reference (nA)

Data Display (nA)

200 200

±1

600 600

±3

1000

1000

±5

Press the [MODE] key to return to normal mode.

3.6 Fieldbus Communication Functional Check

Check for normal function using Fieldbus equipment specified by Yokogawa.

DO202

G

11

12

16

15

13

+

14

17

SH

GALV SENS ●

POL SENS ●

18

+

-

+

-

Decade

Resistance

Box

DC

Generator

2

DC

Generator

1

SW1

Shorting Bar

DC

Source

24VDC

Fieldbus Equipment Specified by Yokogawa

C

R

C

R

R

R

C

C

R

L

L

R=50Ω±1Ω

C=2μF±0.4μF

L=5mH±5%

R

L

=50Ω±1%

SUPPLY

TEMP

+

+

Figure 1 Testing Circuit and Test Equipment