Yokogawa DO202 2-Wire Dissolved Oxygen Analyzer User Manual

Page 47

Advertising
background image

IM 12J05C01-01E

5-7 Parameter Setting

Code Display

Function

Function

detail

X Y Z Default

values

Temperature Measuring Functions
10

*T.SENS

Temp. Comp.

Pt1000 RTD (DO30G)

0

1

22k NTC (Oxyferm)

1

Pb36 = 2k2 NTC (DO30)

2

11

*T.UNIT

Temp units

Celsius (°C)

0

0

Fahrenheit

(°F)

1

12

*T.ADJ

Temp adjust

Adjustment +/- 7.5 °C or +/- 13.5 °F

13

*T.MAN

Manual TC

Automatic temperature comp

0

0

Manual temp compensation

1

25 °C/ 77 °F

14-19

Not

used

Code Display

Function

Function

detail

X Y Z Default

values

Parameter Specific functions
01 *S.TYPE Sensor

type

Galvanic

0 1

Polarographic

1

*V.POL

(Polarographic)

0.675

V

Polarization

Voltage

*SENS Sensor

sensitivity

Polarographic

7.5

nA/ppm

Galvanic

450 nA/ppm (*S.TYPE=0)

04

*SAL.TY

Salinity comp.

Compensation disabled

0

0

Compensation

enabled

1

*[Cl-] NaCl

conc.

Concentration

in g/kg water:

0.0

range 1 to 99.9 g/kg

05-09

Not

used

Code Display

Function

Function

detail

X Y Z Default

values

Calibration Settings

20 *∆T.SEC

Stabilization

Stabilization time range 5- 600 s

60 s

Step

DO step change range 0-50 ppm

∆Z.PPM

0-1999 ppb

0.1 ppm (∆Z.PPM)

∆Z.PPB

1

ppb (∆Z.PPB)

∆Z.SAT

1.0 % SAT (∆Z.SAT)

∆S.PPM

0-600 %SAT

0.1 ppm (∆S.PPM)

(always PPM)

21

*Z.CAL

Zero calibration

Zero calibration disabled

0

0

Zero Calibration enabled

1

*Z.LIM

Zero check limit

Zero check limit (galv.)

0-19.99

μA

19.99

μA

Zero check limit (polar.)

0-199.9

nA

199.9

nA

22

*ZERO

Zero current

Offset range,

0 nA

±Zero Check Limit

*SENS

Sensitivity entry

Sensitivity range 1-1999 nA/ppm

7.5

nA/ppm

450 nA/ppm (*S.TYPE=0)

23-29

Not

used