IM 12J05C01-01E

2-15 General Specifications

2-12. Control Drawing of DO202S FF/PB Specification (FM Intrinsically safe FISCO)

-

+

DO202S-F

or DO202S-P

Transmitter

-

+

Transmitter

Unclassified Location

Classified Location

Division 1

FM Approved

FISCO barrier

Voc (Vt) d17,5 V

Ioc (It) d380 mA

Poc (Pt) d5,32 W

Sensor

Connections

Max. cablelength: 60 mtr.

Cable dia. : 3…12 mm.

FM Class I, DIV. 1, Group ABCD

T4 for ambient temp. d 55 qC

Sensor

Connections

-

+

-

+

FM Approved

Terminator

R = 90..100Ω

C = 0..2,2 μF

FM Approved

Terminator

R = 90..100Ω

C = 0..2,2 μF

x

Sensor(s) are of a passive type to be regarded as 'simple apparatus', devices which

neither store nor generate voltages over 1.5 V, currents over 0.1 A, power over 25 mW or

energy over 20 PJ, or are FM Approvals entity approved and meet connection

requirements.

x

Electrical data of the DO202S-F & DO202S-P:
-

Supply circuit: Vmax = 17.5 V; Imax = 380 mA; Pi = 5.32 W; Ci = 220 pF; Li = 0 μH.

-

Sensor input circuit: Vt = 14.4 V; It = 10 mA; Ca = 643 nF; La = 340 mH

x

Any FM Approved FISCO barrier may be used that meets the following requirements:

Voc or Vt d 17,5 V; Ioc or It d 380 mA; Poc or Pt d 5,32 W

When installing this equipment, follow the manufacturer’s installation drawing.

Installation should be in accordance with ANSI/ISA RP 12.06.01 “Installation of

Intrinsically Safe Systems for Hazardous (Classified) Locations” and the National

Electrical Code (ANSI/NFPA 70).

Associated apparatus connected to the FISCO barrier must not use or generate more

than

250

Vrms or Vdc.

x

Resistance between FISCO Intrinsically Safe Ground and earth ground must be less than

1.0 Ohm.

x

In case of using cable glands in Outdoor location, they shall be UV rated or made of metal.

x

The FISCO concept allows the interconnection of several I.S. apparatus not specifically

examined in such combination. The criterion for such interconnection is that the voltage

(Vmax), the current (Imax) and the power (Pi) which I.S. apparatus can receive and remain

intrinsically safe, considering faults, must be equal to or greater that the voltage (Voc, Vt),

the current (Ioc, It) and the power (Poc, Pt) which can be provided by the FM approved

FISCO barrier. In addition, the maximum unprotected residual capacitance (Ci) and

inductance (Li) of each apparatus (other than the terminator) connected to the Fieldbus

must be less than or equal to 5nF and 10 μH respectively.