2. wiring of power supply, 2-1. general precautions, 2-2. connection of the power supply – Yokogawa DO202 2-Wire Dissolved Oxygen Analyzer User Manual

Page 34: 2-3. switching the instrument on, 2-4. preparation, 2-5. cables, terminals and glands, 3. wiring of sensors, 3-1. wiring the standard galvanic sensor, Warning

Advertising
background image

IM 12J05C01-01E

3-2. Wiring of power supply

3-2-1. General precautions

Do not activate the power supply yet. First make

sure that the DC-power supply is according to the

specifications given.

DO NOT USE ALTERNATING CURRENT OR MAINS

POWER SUPPLY! !

The cable leading to the distributor (power supply) or

safety barrier transports power to and output signal

from the transmitter. Use a two conductor shielded

cable with a size of at least 1.25 mm

2

and an outside

diameter of 6 to 12 mm. The cable gland supplied

with the instrument accepts these diameters. The

maximum length of the cable is 2000 metre, or 1500

metres when using the communications. This ensures

the minimum operating voltage for the instrument.

Grounding:

• If the transmitter is mounted on a grounded

surface (e.g. a metal frame fixed in the soil) the

shield of the 2-wire cable may NOT be connected

to ground at the distributor.

• If the transmitter is mounted on a non-conducting

surface (e.g. a brick wall) it is recommended

to ground the shield of the 2-wire cable at the

distributor end.

3-2-2. Connection of the power supply

The terminal strip is accessed as was described in

section 3-2-1. Use the left-hand gland to insert the

supply/output cable to the transmitter. Connect the

supply to the terminals marked +, - and G as is indi-

cated in figures 3-7.

3-2-3. Switching the instrument on

After all connections are made and checked, the

power can be switched on from the distributor.

Observe the correct activation of the instrument at

the display. If for any reason the display does not

indicate a value, consult the trouble shooting section.

3-2-4. Preparation

Refer to figure 3-4. The power/output connections

and the sensor connections should be made. Refer

to figure 3-7. The terminals are of a plug in style for

ease of mounting.

To open the EXA 202 for wiring:

1. Loosen the four frontplate screws and remove

the cover.

2. The terminal strip is now visible.

3. Connect the power supply. Use the gland on the

left for this cable.

4. Connect the sensor input, using the gland on

the right (see fig. 3-5). Switch on the power.

Commission the instrument as required or use

the default settings.

5. Replace the cover and secure frontplate with the

four screws.

6. Connect the grounding terminals to protective

earth.

7. The optional hose connection is used to guide the

cables coming from an immersion fitting through a

protective plastic tubing to the transmitter.

3-2-5. Cables, terminals and glands

The DO202 is equipped with terminals suitable for

the connection of finished cables in the size range:

0.13 to 2.5 mm (26 to 14 AWG). The glands will

form a tight seal on cables with an outside diameter

in the range of 6 to 12 mm (0.24 to 0.47 inches).

3-3. Wiring of sensors

3-3-1. Wiring the standard galvanic sensor

The sensor cable has markers on the individual wires.

These markings refer to the markers on the terminals.

The temperature sensor has two wires with the

markings T1, T2 and/or 11, 12 and must be con-

nected to terminal 11 and 12.

The measuring electrode: the cathode is marked with

IE and/or 13 and must be connected to terminal 13.

The reference electrode: the anode is marked with

RE and/or 15 and must be connected to terminal 15.

The liquid earth (solution ground) is marked with 16

and must be connected to terminal 16.

The overall shield of the cable is marked with 14

and must be connected to terminal 14.

Place the jumper in the GALVANIC position.

WARNING

WARNING

3-3 Installation and Wiring

Grounding terminal

Power/Output

cable gland

Sensor

cable gland

Figure 3-5. Glands to be used for cabling

mA

mA

mA

Advertising