Caution – Yokogawa DO202 2-Wire Dissolved Oxygen Analyzer User Manual

Page 90

Advertising
background image

IM 12J05C01-01E

2-2 General Specifications

2-2. Performance specifications
A. Performance in ppm mode

- Linearity

: ±0.05 ppm or ±0.8% FS,

whichever is greater

- Repeatability

: ±0.05 ppm or ±0.8% FS,

whichever is greater

- Accuracy

: ±0.05 ppm or ±0.8% FS,

whichever is greater

B. Performance in ppb mode

- Linearity

: ±1 ppb or ±0.8% FS,

whichever is greater

- Repeatability

: ±1 ppb or ±0.8% FS,

whichever is greater

- Accuracy

: ±1 ppb or ±0.8% FS,

whichever is greater

C. Performance in Temperature

- Linearity

: ±0.3ºC

- Repeatability

: ±

0.1ºC

- Accuracy

: ±0.3ºC

Note on performance specifications:

The specifications are at reference temperature

and with simulated inputs, because the

DO202G can be used with many different

sensors with their unique characteristics.

The following tolerance is added to above

performance.

mA output tolerance : ± 0.02 mA of

"4 - 20 mA"

D. Ambient temperature

Operating temperature

-10 to 55ºC

(10 to 130ºF)

Storage temperature

-30 to 70ºC

(-20 to 160ºF)

E. Humidity

10 to 90% RH non-condensing

F. Housing

Cast aluminum housing with chemically

resistant coating, cover with flexible

polycarbonate window. Case color is off-white

with moss green cover. Cable entry is through

two polyamide cable glands. Cable glands are

provides for wiring up to 2.5 mm.

G. Environmental protection

IP65, NEMA 4X.

H. Mounting

Pipe, wall or panel mounting using optional

hardware.

I. Data protection

EEPROM for configuration; Lithium battery for

clock.

J. Watchdog timer: Checks microprocessor.
K. Autoreturn

The transmitter returns into measuring mode

automatically when no keystroke is made for 10

minutes.

L. Operation protection

All three levels can be protected by a user

programmable 3 digit password

M. EMC Conformity standards

,

EN 61326-1 Class A, Table 2

(For use in industrial locations)

EN 61326-2-3

EN 61326-2-5 (Profibus communication may be

influenced by strong electromagnetic field.)

CAUTION

This instrument is a Class A product, and it is

designed for use in the industrial environment.

Please use this instrument in the industrial

environment only.

N. Explosionproof type

Refer to Control Drawings.

Item

Description

Code

Factory

Mutual (FM)

-N

-A

-N

-A

CENELEC

ATEX

2.EPS

FM Non-incendive safe Approval

Applicable standard: FM3600, FM3611, FM3810

Non-incendive Safe for Class I, Division 2,

Groups ABCD, Zone 2

Temp. Class: T4, Amb. Temp.: -10 to 55°C

Non-incendive Safe Apparatus Parameters

Vmax=31.5 V, Ci=22 nF, Li=35 μH

FM Intrinsically safe Approval

Applicable standard: FM3600, FM3610, FM3810

Intrinsically Safe for Class I, Division 1, Groups ABCD

Class I, Zone 0, AEx ia IIC

Temp. Class: T4, Amb. Temp.: -10 to 55°C

Intrinsically Safe Apparatus Parameters

Vmax=31.5 V, Imax=100 mA,

Pmax=1.2 W, Ci=22 nF, Li=35 μH

CENELEC ATEX (KEMA)

Type of protection "n"

Applicable standard: EN60079-0:2006,

EN60079-15:2003

Certificate: KEMA

06ATEX0225

EEx nA [nL] IIC, Group: II, Category: 3G

Temp. Class: T4, Amb. Temp.: -10 to 55°C

T6, Amb. Temp.: -10 to 40°C

Ui=31.5 V, Ci=22 nF, Li=35 μH

CENELEC ATEX (KEMA) Intrinsically safe Approval

Applicable standard: EN60079-0, EN50020

EN60079-26

Certificate: KEMA

06ATEX0224 X

Ex ia IIC, Group: II, Category: 1G

Temp. Class: T4, Amb. Temp.: -10 to 55°C

T6, Amb. Temp.: -10 to 40°C

Ui=31.5 V, Ii=100 mA, Pi=1.2 W, Ci=22 nF, Li=35 μH

Item

Description

Code

Factory

Mutual (FM)

FM Non-incendive safe Approval

Applicable standard: FM3600, FM3611, FM3810

Non-incendive Safe for Class I, Division 2,

Groups ABCD, Zone 2

Temp. Class: T4, Amb. Temp.: -10 to 55°C

Non-incendive Safe Apparatus Parameters

FM Intrinsically safe Approval

Applicable standard: FM3600, FM3610, FM3810

Intrinsically Safe for Class I, Division 1, Groups ABCD

Class I, Zone 0, AEx ia IIC

Temp. Class: T4, Amb. Temp.: -10 to 55°C

Intrinsically Safe Apparatus Parameters

FM.EPS

Vmax=32 V, Pmax=1.2 W,

Ci=220 pF, Li=0 μH

Vmax=32 V, Pmax=5.32 W,

Ci=220 pF, Li=0 μH

Entity

FNICO

-F

-P

or

-D

-B

or

Vmax=24 V, Imax=250 mA,

Pmax=1.2 W, Ci=220 pF, Li=0 μH

Vmax=17.5 V, Imax=380 mA,

Pmax=5.32 W, Ci=220pF, Li=0 μH

Entity

FISCO

mA

Advertising