3-7. installation in:hazardous area-non-incendive – Yokogawa DO202 2-Wire Dissolved Oxygen Analyzer User Manual

Page 36

Advertising
background image

IM 12J05C01-01E

3-5 Installation and Wiring

Extension cable may be purchased in bulk quanti-

ties, cut to length. Then it is necessary to terminate

the cable as shown below.

Termination procedure for WF10 cable.

1. Slide 3 cm of heat shrink tube (9 x 1.5) over the

cable end to be terminated.

2. Strip 9 cm of the outer (black) insulating material,

taking care not to cut or damage internal cores.

Figure 3-9a.

3. Remove loose copper screening, and cut off the

cotton packing threads as short as possible.

4. Strip insulation from the last 3 cm of the brown,

and the white coaxial cores.

Figure 3-9b.

5. Extract the coaxial cores from the braid, and trim

off the black (low-noise) screening material as

short as possible.

6. Insulate the overall screen and drain wire (14) and

the 2 coaxial screens with suitable plastic tubing.

7. Strip and terminate all ends with suitable (crimp)

terminals and identify with numbers as shown.

Figure 3-9c.

8. Finally shrink the overall heat shrink tube into

position.

3 cm

heat shrink

9 cm

remove insulation

3 cm

cotton threads

11

12
14

15
16

13

Red

Black

Blue

White

Brown

3-3-6. Additional precautions for installations in

hazardous areas

Make sure that the total of capacitance and induct-

ances connected to the input terminals of the EXA

DO202S do not exceed the limits given in the certifi-

cate.

This sets a limit to the cable and extensions used.

– The intrinsic safe version of the DO202S

instrument can be mounted in Zone 1 (DO202S-B,

-N, -D).

– The sensors can be installed in Zone 0 or Zone 1

if a safety barrier according to the limits given in

the system certificate is used.

– Ensure that the total of capacitance and

inductances connected to the terminals of the

EXA DO202S do not exceed the limits given in the

certificate of the safety barrier or distributor.

– The cable used should preferably have a BLUE

colour or marking on the outside.

– Installation for (sensors in Zone 0 or 1):

Generally, the distributor with input/output isolation

has no external earth connection. If there is

an earth connection on the distributor and the

external connection of the transmitter is connected

to “protective” earth, the shield of the 2-wire cable

may NOT be connected to “protective” earth at the

distributor too.

3-3-7. Installation

in:

Hazardous Area-Non-Incendive

The EXA DO202S-B, -N, -D may be installed in a

Category 3, Zone 2, Div.2 area without the use of

safety barriers. Maximum permissible supply voltage

31.5V.