IM 12J05C01-01E

Error Generation

Error content and Causes

Measures

No. Mode

E10 All modes

EEPROM writing failure

Turn off the power and then turn it

Electronic circuit failure

back on and check whether or not the

system returns to normal.

If the failure occurs again, contact

Yokogawa to request repair.

E15 CODE 11

Temperature error correction failure

Accurately measure the temperature of

The difference between the corrected value

the measuring solution (dissolved oxygen

and the standard value exceeds the range

sensor).

corresponding to ±7.5°C (±13 5°F).

Examine whether the sensor cable fails

• Temperature entry is not accurate.

(partial disconnection or degraded

• Sensor cable failure

insulation) and then correct the

temperature error again.

E16 All modes

Call for maintenance-interval time exceeded

Perform maintenance. Reset interval.

E17 *OUTP

Output span too small.

Re-enter a value within the permissible range.

CODE 35

Min. span is 1ppm, 25ppb or 10%

E18 CODE 35

Table values make no sense.

Reprogram.

E19 At data entry

Configuration failure

Re-enter a value within the permissible

A value not within the permissible range is entered.

range.

E20 All modes

Instrument initialization failure

Contact Yokogawa

E21 All modes

EPROM comparison failure

Contact Yokogawa

Troubleshooting 8-2

Note:

It is possible that the error message does

not disappear immediately after corrective

measures.

Then turn power off and the turn it back on

and check whether the system returns to

normal.