IM 12J05C01-01E

Parameter Setting 5-10

5-3-7. General

Code 70

*LOAD

The load defaults code allows the instrument to be returned to the default set up

with a single operation. This can be useful when wanting to change from one

application to another.

5-3-8. Test and setup mode

Code 80

*TEST

Not used

Note:

attempting to change data in service code 80 and above without the proper instructions and

equipment, can result in corruption of the instrument setup, and will impair the performance of the

unit.

5-3-6. Communication setup

Code 60 *COMM. The settings should be adjusted to suit the communicating device connected

to the output. The communication can be set to HART®.

*ADDR.

Select address 00 for point to point communication with 4-20mA transmission.

Addresses 01 to 15 are used in multi-drop configuration (fixed 4mA output).

Code 61 *HOUR

The clock/calendar for the logbook is set for current date and time as reference.

*MINUT

*SECND

*YEAR

*MONTH

*DAY

Code 62

*ERASE

Erase logbook function to clear the recorded data for a fresh start. This may be

desirable when re-commissioning an instrument that has been out of service for a

while.