4-1. Operator interface

This section provides a survey of the operation of

the EXA operator interface. The basic procedures for

obtaining access to the three levels of operation are

described briefly. For a step-by-step guide to data

entry, refer to the relevant section of this user’s manu-

al. Figure 4-1 shows the EXA operator interface.

LEVEL 1: Maintenance

These functions are accessible by pushbutton

through a flexible front cover window. The functions

make up the normal day-to-day operations that an

operator may be required to complete. Adjustment

of the display and routine calibration are among the

features accessible in this way. (See table 4-1).

LEVEL 2: Commissioning

A second menu is exposed when the EXA front

cover is removed and the display board is revealed.

Users gain access to this menu by pressing the but-

ton marked * in the lower right of the display board.

This menu gives access to the service menu.

(See table 4-1).

LEVEL 3: Service

For more advanced configuration selections, press

the button marked *. Now push the “YES” button.

Selecting and entering “Service Code” numbers in

the commissioning menu provide access to the

more advanced functions. An explanation of the

Service Codes is listed in chapter 5 and an over-

view table is shown in chapter 11.

Operation: Display Functions and Settings 4-1

4. OPERATION; DISPLAY FUNCTIONS AND SETTING

Routine

Function

Chapter

Maintenance

CALIB

AIR.CAL

H2O.CAL

SENS

ZERO

Calibration in air (preferred method)

Calibration in aerated water

Manual sensitivity calibration

Manual zero calibration

6

6

6

6

DISP

M.TEMP

M.PRESS

Read auxiliary data or set message display

Adjust manual temperature

Adjust manual pressure

4

5

5

Commissioning

OUTP

HOLD

Adjust the output range

Activate the hold function

5

5

Service
(access to coded entries from
the commissioning level)

SERV

Fine tune the specialized functions of the transmitter

5

mA

Note

: All three levels may be separately protected by a password. See Service Code 52 in chapter 5

Service Code table for details on setting passwords.

Table 4-1. Operations overview