1-2 Appendix

A-1-2 Configuration Checklist for DO202

Standard configuration

Options

Reference for change

Measured variable(s)
Primary inputs

DO and Temp

DO units

ppm

ppb. % saturation

code 56

Temperature unit

Celsius

Fahrenheit

code 11

mA Outputs
Analog output

4-20mA for ppm

ppb or %sat

code 56, 31

Output linearization

disabled

ppm, ppb or %sat

code 35

Burn out

disabled

burn low (3.6) / high (21) on mA output code 32

Communication
Digital interface

HART

disabled

code 60

Communication software external

HHC

contact factory

Interface

Variables on display

ppm and temp

%sat, °C, SE, ZR, REL

“display”

Password protection

disabled

for mains/ comm./ serv level

code 52

Autoreturn

return to measure in 10 minutes

enable or disable

code 50

Add. functions in MAINT disabled

manual pressure

code 51

Diagnostics
Check on Slope

active

1-1999 na/ppm

code 22

Check on zero

disabled

enable or disable

code 02, 21, 22

Check on stability

0.1 ppm per 60 s

0-50 ppm in 5-600 s

code 20

Compatibility
DO sensor

SM31

DO30G

code 01, 10

Temperature sensor

22k NTC

Pt 1000

code 10

Sensor principle

polarographic

galvanic

code 01, “wiring”

Manual temp. comp.

disabled

disable or enable

code 13

Special Features
Salinity compensation

disabled

comp. for 1-99.9 ppt NaCI

code 04

Temperature calibration none

adjustment ±7.5 °C or ±13.5°F

code 12

Zero calibration

disabled

disable or enable

code 21

Atm. press. compensation manual

Soft fail alarm

disabled

possible for E1..E2, E7..E9, E16

code 53

Logbook

disabled

2 volumes of 50 events

code 61,62

mA

mA

Through HART the user has access to the process

variables:

PV: User selected process value: ppm, ppb or %

saturation

SV: Temperature

TV: % Saturation

QV: sensor current

Each of these process variables can be decoded into

current output or contact output through the HIM: Hart

Interface Module. Yokogawa offers standard a HIM

with two 0- 20 or 4- 20 mA outputs and three contact

outputs with 24 VDC power supply.

Full communication with the DO202 can be done

through various brands of HART Handheld termi-

nals (HHT) or through Pactware. This program and

the Yokogawa DTM library can be downloaded

from YOKOGAWA WEB sites free of charge by

Yokogawa customers.

Yokogawa recommends the use of MH-02 modem

for HART-RS232 conversion.

A-1-3 HART process variables