Calibration procedure, 1. general, 1-1. calibration methods – Yokogawa DO202 2-Wire Dissolved Oxygen Analyzer User Manual

Page 52

Advertising
background image

IM 12J05C01-01E

Calibration Procedure 6-1

6. CALIBRATION PROCEDURE

Calibration of the dissolved oxygen analyzer is per-

formed in the following situations:

• When a new dissolved oxygen sensor is installed.

• When the membrane is replaced and/or the

electrolyte solution is replaced

• When the sensor has been disassembled and

reassembled for maintenance

• When the measuring error after cleaning exceeds

the acceptable deviation from reference method

• When error message E16 indicates the need for

maintenance

6-1. General

6-1-1. Calibration methods

There are four methods for calibration of the EXA

dissolved oxygen analyzer: air calibration, water

calibration, sensitivity and zero (offset) calibration.

Air calibration

This is the most common and easiest way to cali-

brate the analyzing system. The calibration is done

in AIR.CAL mode in maintenance level.

Water calibration

This is the traditional way to calibrate dissolved oxy-

gen analyzers in laboratory environment. . Accurate

results can be obtained if the calibration is per-

formed carefully. The water calibration may include

a zero calibration if so selected in service code 21.

Normally only a zero check is performed without

calibration, since a good sensor has no zero signal.

Water calibration is not possible in salt water unless

salinity compensation is activated in service code

04. The calibration is done in H2O.CAL mode in

maintenance level

Note:

In normal industrial environments better

results are obtained with air calibration, which

is therefore the recommended method.

Manual calibration

This is a calibration method whereby the sensor is

not exposed to a calibration fluid, but is calibrated

by comparison with a reference method. This refer-

ence method is normally a laboratory analysis

of the water sample: either by Iodometric titration

according ISO 5813 or an electrochemical method

according ISO 5814.

The calibration is done in SENS/ZERO mode in

maintenance level.

6-1-2. Diagnostic functions performed during

calibration

The calibration is a semi-automatic calibration,

which means that the sensor output is used for

calculation of sensor parameters after the read-

ings have stabilized. The criteria for stabilization

are set in service code 20. If stable readings are

not achieved within one hour error E1 message will

appear on the display and the procedure is aborted.

The calculated sensor parameters are compared

with preset tolerances and if these conditions are

not met, Error E2 (Zero Error) or E3 (Sensitivity

Error) will be signalled.

Notes:

1.

Error messages that appear initially indicate

inadequate settings of the diagnostic variables

in service codes 01, 20 and/or 22.

2. Patience improves the accuracy of calibration.

When the EXA prompts the user that the

preset criteria are met by showing CAL.END

message it only means that the minimum

criteria are met. When the signal still changes

after this initial acceptance, EXA will

automatically update the sensor characteristics

with the new information.

Advertising