3-4. ma output functions, 3-5. user interface – Yokogawa DO202 2-Wire Dissolved Oxygen Analyzer User Manual

Page 48

Advertising
background image

IM 12J05C01-01E

Parameter Setting 5-8

5-3-4. mA output functions

Code 31

*OUTP.F For the DO202 the output may be chosen as linear to input, or configured in a

21 point table to a particular linearization. Enable the table setup in code 31, and

configure the table in code 35.

Code 32

*BURN

Diagnostic error messages can signal a problem by sending the output signals

upscale (21 mA) or downscale (3.6 mA when HART comm. non-used, 3.9 mA

when HART comm. used). This is called upscale or downscale burnout, from

the analogy with thermocouple failure signaling of a burned-out or open circuit

sensor. The pulse burnout setting gives a 21 mA signal for the first 30 seconds

of an alarm condition. After the “pulse” the signal returns to normal. This allows a

latching alarm unit to record the error. In the case of the EXA the diagnostics are

extensive and cover the whole range of possible sensor faults.

Note:

Burnout downscale can only be selected if the HART communication is turned of in service code 60.

Code 35

*TABLE

The table function allows the configuration of an output curve by 21 steps

(intervals of 5%). The following example shows how the table may be configured

to linearize the output with a mA curve.

5-3-5. User interface

Code 50

*RET

When Auto return is enabled, the transmitter reverts to the measuring mode from

anywhere in the configuration menus, when no button is pressed during the set

time interval of 10 minutes.

Code 51

*P.MAN

In Maintenance mode the pressure can be entered (M. PRESS).

*P.%

*P.CAL

Code 52

*PASS

Passcodes can be set on any or all of the access levels, to restrict access to the

instrument configuration.

Code 53

*Err. xx

Error message configuration. Two different types of failure mode can be set.

Hard fail gives a steady FAIL flag in the display.

Soft fail gives a flashing FAIL flag in the display. The call for maintenance is

a good example of where a SOFT fail is useful. A warning that the regular

maintenance is due, should not be used to shut down the whole measurement.

Code 55

*CALL.M

Call for maintenance is a contact trigger to signal that the system has been in

service for longer than the set time without calibration. The user can set up to

250 days as a routine service interval.

Code 56

*UNIT

Select ppm, ppb or % saturation as the main unit, to suit the application.

mA