Yokogawa DO202 2-Wire Dissolved Oxygen Analyzer User Manual

Page 81

Advertising
background image

IM 12J05C01-01E

2-16 Appendix

3/3

QIS 12J05C01-71E

i. Press

the

[ ] key. The instrument returns to the measurement mode.

Set the display to nA current indication by following the steps below.

a. Press the [MODE] key. The message display will show “CALIB.”

b. Press the [NO] key once. The message display will show “DISP.”

c. Press the [YES] key once. The message display will show temperature indication.

d. Press the [NO] key five times. The message display will show nA current indication.

Change the current of DC generator 2 as shown in Table 3. The corresponding indication must

be within the range.

Table 3

Reference (nA)

Data Display (nA)

200 200

±1

600 600

±3

1000

1000

±5

Press the [MODE] key to return to normal mode.

3.6 Profibus

Communication

Functional

Check

Check for normal function using Profibus equipment specified by Yokogawa.

In the tests of Item 3.3 to 3.5, check the communication function using Profibus communication.

DO202

G

11

12

16

15

13

+

14

17

SH

GALV SENS ●

POL SENS ●

18

+

-

+

-

Decade

Resistance

Box

DC

Generator

2

DC

Generator

1

SW1

Shorting Bar

Profidbus Equipment Specified by Yokogawa

C

R

C

R

SUPPLY

TEMP

+

R=50

Ω± Ω

C=2μF±0.4μF

+

DA/PA Coupler

DC

Source

24VDC

Figure 1 Testing Circuit and Test Equipment