In this state, change the resistance value of the decade resistance box as shown in Table 1.

The corresponding temperature indication must be within the range.

Table 1

Reference

Temperature (°C)

Resistance of

Resistance Box (Ω)

Data Display (°C)

–9 99.6k

–9

±0.3

25 22k

25

±0.3

140 653.8

140

±0.3

3.4 Galvanic

Indication

Check

Turn on SW1.

Set the sensor type to galvanic.

Setting Service Code 01 to 0.

a. Press the [ ] key. The message display will show “ SERV.”

b. Press the [YES] key. The data display will show “00” with the first digit of 0 flashing.

c. Press the [>] key once. The second digit of 0 flashes.

d. Press the [

š

] key once. The data display will show “01” with 1 flashing.

e. Press the [ENT] key. The message display will show “ STYPE.”

f. Press

the

[

š

] key once. The data display will change from “1” to “0.”

g. Press the [ENT] key until the message display shows “ SERV.”

h. Press the [ ] key. The instrument returns to the measurement mode.

Set the display to μA current indication by following the steps below.

a. Press the [MODE] key. The message display will show “CALIB.”

b. Press the [NO] key once. The message display will show “DISP.”

c. Press the [YES] key once. The message display will show temperature indication.

d. Press the [NO] key five times. The message display will show μA current indication.

Change the current of DC generator 1 as shown in Table 2. The corresponding indication must

be within the range.

Table 2

Reference (μA)

Data Display (μA)

10 10

±0.06

30 30

±0.10

50 50

±0.20

Press the [MODE] key to return to normal mode.

3.5 Polarographic

Indication

Check

Turn off SW1.

Set the sensor type to polarograph.

Setting Service Code 01 to 1.

a. Press the [ ] key. The message display will show “ SERV.”

b. Press the [NO] key twice.

c. Press the [YES] key. The data display will show “00” with the first digit of 0 flashing.

d. Press the [>] key once. The second digit of 0 flashes.

e. Press the [

š

] key once. The data display will show “01” with 1 flashing.

f. Press the [ENT] key. The message display will show “ STYPE.”

g. Press the [

š

] key once. The data display will change from “0” to “1.”

h. Press the [ENT] key until the message display shows “ SERV.”