3-1. preparation, 3-3. procedure for water calibration – Yokogawa DO202 2-Wire Dissolved Oxygen Analyzer User Manual

Page 54

Advertising
background image

IM 12J05C01-01E

Calibration Procedure 6-3

6-3-3. Procedure for water calibration

NO

SC21=0

NO

SC21=1

NO

NO

YES

NO

YES

NO

NO

YES

YES

Press the MODE key. CALIB appears in the display, and the

YES/NO

key prompt flags flash. Press YES.

NO

YES

NO

YES

YES

NO

After briefly displaying WAIT,

the display returns to the

normal readout.

YES

Sensor can be returned

to process water.

Press YES for single point

adjustment.

YES

NO

NO

YES

NO

NO

YES

YES

Go to

6-4-2

Go to

6-4-3

NO

YES

The instruments

waits for the

reading to

stabilise.

(The display

flashes). When

reading Is stable,

the CAL END

message

appears.

ENT

FAIL

TEMP.MAN

YES NO

ENT

NO

MODE

YES

MODE

%sat

ppb

CAL
DISPLAY

ppm

HOLD

OUTPUT

SET HOLD

SERVICE

YES

NO

YES

YES

YES

YES

YES

YES

to *1

NO

Note: To start calibration with a zero solution

(zero-dissolved-oxygen solution.)

from *1

NO

NO

to *1

6-3. Calibration procedure using water calibra-

tion method

6-3-1. Preparation

Move the sensor to a maintenance site and wash

off any dirt on the membrane. Lightly wipe off any

remaining water from the membrane with a soft tissue.

Prepare the necessary equipment and reagents to be

used for the span and (if required) zero calibration.

Equipment for span calibration includes:

• Beaker or bucket

• Magnetic stirrer or other means of agitating the water

• Demineralized water or salty process water if

salinity compensation is activated

• Air supply pump (aquarium pump)

• Glass diffuser to generate small air bubbles

Note:

It takes 15 to 30 minutes of aeration before

it can be assumed that the water is fully

saturated with air.

Reagent for zero calibration:

• Sodium sulfite

• demineralized water

The zero oxygen solution contains 20- 30 grams of

sodium sulfite per liter.

6-3-2. Calibration

operation

(water calibration method)

If both zero and span calibration are used, it is bet-

ter to start with span calibration in spite of normal

calibration sequence with analog instrumentation.

Otherwise the sensor must be rinsed off carefully to

prevent ingress of oxygen absorbing sulfite in the

aerated water.

Note:

Zero calibration in the sodium sulfite solution

takes a long time. Even after CAL.END

prompt it is necessary to wait 40-50 minutes

before zero adjustment can be made.

Otherwise calibration errors are introduced in

the measurement. It is recommended not to

do a zero calibration, but a zero check only.

A good clean sensor does not have a zero

current and does not require zero calibration.

If the sensor has a substantial zero current,

the sensor has to be disassembled, cleaned

and filled with fresh electrolyte solution and a

new membrane has to be fitted.

Calibration may take longer than 10 minutes.

Before starting the calibration procedure,

check the setting of Auto Return (*RET.) in

Service Code 50. It is recommended to set

the Auto Return to “0” (No return to meas.

from HOLD). After the calibration procedure

has been completed, set the Auto Return to

the desired setting in Service Code 50.

Advertising