4-2 Operation: Display Functions and Settings

ENT

FAIL

TEMP.MAN

YES NO

ENT

NO

MODE

YES

MODE

%sat

ppb

CAL
DISPLAY

ppm

HOLD

OUTPUT

SET HOLD

SERVICE

Fail flag

Commissioning
function menu

Commissioning mode
access key

Measure/Maintenance
mode key

Broken line indicates area that can be
seen through front cover

Adjustment keys
> :

Choose digit to
adjust

^

: Adjust digit

ENT : Confirm change

Selection keys
YES : Accept setting
NO : Change setting

Key prompt flags

Message display

Main display

Menu pointer flags

Figure 4-1. DO202 operator interface

4-2. Explanation of operating keys

MODE key

This key toggles between the measuring and

maintenance modes. Press once to obtain

access to the maintenance function menu.

CALIB

DISP

Press again to return to the measuring mode.

YES/NO keys

These are used to select choices from the menu.

YES is used to accept a menu selection.

NO is used to reject a selection, or to move

ahead to the next option.

DATA ENTRY keys (

)

Is used as a “cursor” key. Each press

on this key moves the cursor or flashing

digit one place to the right. This is used

to select the digit to be changed when

entering numerical data.

Is used to change the value of a selected

digit. Each press on this key increases the

value by one unit. The value can not be

decreased, so in order to obtain a lower

value, increase past nine to zero, then

increase to the required number.

When the required value has been set

using the

and

keys, press

to

confirm the data entry. Please note that

the EXA does not register any change of

data until the ENT key is pressed.

This is the commissioning mode key. It is

used to obtain access to the commissio-

ning menu. This can only be done with the

cover removed or opened. Once this button

has been used to initiate the commissioning

menu, follow the prompts and use the other

keys as described above.

4-3. Passcode protection

In Service Code 52, passcode protection can be set

for each one of the three operating levels, individu-

ally. This procedure should be completed after the

initial commissioning (setup) of the instrument. The

passcodes should then be recorded safely for future

reference.

When passcodes have been set, the following addi-

tional steps are introduced to the configuration and

programming operations:

Maintenance

Press MODE key. The display shows 000 and *PASS*

Enter a 3-digit passcode as set in Service Code 52

to obtain access to the Maintenance Mode.

Commissioning

Press key. The display shows 000 and *PASS*

Enter a 3-digit passcode as set in Service Code 52

to obtain access to the Commissioning Mode.

Service

From the Commissioning menu, select *Service by

pressing YES key. The display shows 000 and *PASS*

Enter a 3-digit passcode as set in Service Code 52

to obtain access to the Service Mode.

Note:

See Service Code 52 for the setting of

passcodes.