Yokogawa DO202 2-Wire Dissolved Oxygen Analyzer User Manual

Page 49

Advertising
background image

IM 12J05C01-01E

5-9 Parameter Setting

Code Display

Function

Function

detail

X Y Z Default

values

Output Functions
30

Not

used

31

*OUTP.F

mA output functions

Linear

0

0

Linear

Table

1

32

*BURN

Burn function

No burnout

0

0

No Burn.

Burnout

downscale

1

Burnout

upscale

2

Pulse

burnout

3

33, 34

Not used

35

*TABLE

Output table for mA

*0%

Linearization table for mA in 5% steps.

*5%

The measured value is set in the main

*10%

display using the >, ^, ENT keys, for

...

each of the 5% interval steps.

...

Where a value is not known, that value may

*95%

be skipped, and a linear interpolation will

*100%

take

place.

36-39

Not

used

Code Display Function

Function

detail

X

Y

Z

Default

values

User Interface
50

*RET

Auto return

No return to meas. from HOLD

0

1

Return to meas after 10 min

1

51

*P.MAN

Manual pressure No manual pressure in Maint.mode

0

0

Manual pressure in Maint.mode

1

*P.%

Enter Manual process pressure 0-999.9 kPa

101.3 kPa

*P.CAL

Enter Cal. pressure in Maint.mode 0-999.9 kPa

101.3 kPa

52

*PASS

Password

Direct access to MAINT

0

0.0.0

Pass: 111 for #=1, MAINT protected by password

#

333 for #=2,

Direct access to COMM

0

777 for #=3,

COMM protected by password

#

888 for #=4,

Direct access to SERV

0

123 for #=5,

SERV protected by password

#

957 for #=6,

331 for #=7,

546 for #=8 and

847 for #=9

53

*Err.01

Error setting

Calibration not stable

Soft fail

0

Calibration not stable

Hard fail

1

1

Hard

*Err.02

Zero out of limits

Soft fail

0

Zero out of limits

Hard fail

1

1

Hard

*Err.07

Temperature sensor open

Soft fail

0

Temperature sensor open

Hard fail

1

1

Hard

*Err.08

Temp. sensor shorted

Soft fail

0

Temp. sensor shorted

Hard fail

1

1

Hard

*Err.09

Cell current abnormal

Soft fail

0

Cell current abnormal

Hard fail

1

1

Hard

*Err.16

Call for maintenance

Soft fail

0

0

Soft

Interval time exceeded

Call for maintenance

Hard fail

1

Interval time exceeded

55

*CALL.M E16 setting

No call for maintenance

0

0

Call for maint. enabled

1

*CALL .M

E16 message after 1- 250 days

250

56

*UNIT

PV unit

Unit of measurement ppm

0

0

Unit of measurement ppb

1

Unit of measurement % sat.

2

57-59

Not

used

mA