Installation and wiring, 1. installation and dimensions, 1-1. installation site – Yokogawa DO202 2-Wire Dissolved Oxygen Analyzer User Manual

Page 32: 1-2. mounting methods

Advertising
background image

IM 12J05C01-01E

3-1. Installation and dimensions

3-1-1. Installation site

The EXA transmitter is weatherproof and can be installed

inside or outside. It should, however, be installed as

close as possible to the sensor to avoid long cable runs

between sensor and transmitter. In any case, the cable

length should not exceed

30 mtr (100 feet).

Select an installation site where:

• Mechanical vibrations and shocks are negligible

• No relay/power switches or RF transmitters are in

the direct environment

• Access is possible to the cable glands (figure 3-1)

• The transmitter is not mounted in direct sunlight or

severe weather conditions

• Maintenance procedures are possible (avoiding

corrosive environments)

The ambient temperature and humidity of the instal-

lation environment must be within the limits of the

instrument specifications. (See chapter 2).

When the instrument with Suffix Code "-B,-N,-D" is

used, take measures so that the display window is

not exposed to direct sunlight.

3-1-2. Mounting methods

Refer to figures 3-2 and 3-3. Note that the EXA

transmitter has universal mounting capabilities:

• Panel mounting using optinal bracket, refer to Fig.

3-2a.

• Panel mounting using two (2) self-tapping screws,

refer to Fig. 3-2b.

• Surface mounting on a plate

(using bolts from the back)

• Wall mounting on a bracket

(for example, on a solid wall)

• Pipe mounting using a bracket on a horizontal or

vertical pipe (Nominal pipe diameter 50 A)

3. INSTALLATION AND WIRING

Figure 3-1. Housing dimensions and layout of glands

3-1 Installation and Wiring

162

202

155

180

30

77

50

42

68

115

130

56±0.2

Hood (Option)
Option code : / H

Grounding terminal

(M4 screw)

Transmission signal cable inlet

Cable gland : Pg13.5

Sensor cable inlet

cable gland (Pg13.5)

M6 screw (32 )

1.eps

9

34

(7.95)

(6.4)

(6.1)

(7)

(2)

(5.1)

(1.2)

(3)

(4.5)

(1.65)

(2.7)

(0.35)

(1.26) depth

(1.3)

(2.20)

1 to 10

205

185

3.9

80

38

60

30

Panel thickness

Panel mounting

bracket

173

+1.1

0

156

+1.1

0

121

138

PANEL CUTOUT

5.eps

(0.04 to 0.39)

(7.28)

(8.07)

(1.5) (3.15) (0.15)

(6.14)

(4.76)

(6.81) (5.43)

(1.2)

(2.36)

49

Adapter

G1/2 screw (/AFTG)

1/2 NPT screw (/ANSI)

Approx.

55

2.eps

(2.2)

(1.93)

Unit: mm (inch)

Figure 3-2a. Panel mounting diagram