Yokogawa DO202 2-Wire Dissolved Oxygen Analyzer User Manual

Page 72

Advertising
background image

IM 12J05C01-01E

Appendix 2-7

3/4

QIS 12J05C01-21E

3.5 Galvanic

Indication

Check

Following Section 3.4, press the [ENT] key until the message display shows “GALVA.” Then

turn SW1 off. Change the current of the DC generator 1 and check the data display. The value

on the data display must be within the range shown in Table 3.

Table 3

Reference (μA)

Data Display (μA)

10 10

±0.06

30 30

±0.1

50 50

±0.2

3.6 Polarograph

Indication

Check

Press the [ENT] key until the message display shows “POLAR.1.” Then turn SW1 on. Change

the current of the DC generator 2 and check the data display. The value on the data display

must be within the range shown in Table 4.

Table 4

Reference (nA)

Data Display (nA)

2 2

±0.02

6 6

±0.04

12 12

±0.07

Press the [ENT] key. The message display will show “POLAR.2.” Change the current of the DC

generator 2 and check the data display. The value on the data display must be within the range

shown in Table 5.

Table 5

Reference (nA)

Data Display (nA)

20 20

±0.1

60 60

±0.3

100 100

±0.5

Press the [ENT] key. The message display will show “POLAR.3.” Change the current of the DC

generator 2 and check the data display. The value on the data display must be within the range

shown in Table 6.

Table 6

Reference (nA)

Data Display (nA)

200 200

±1

600 600

±3

1000

1000

±5

Press the [ENT] key until the message display shows “READY.” Press the [ENT] key again to

finish the tests.