Yokogawa DO202 2-Wire Dissolved Oxygen Analyzer User Manual

Page 67

Advertising
background image

IM 12J05C01-01E

2-2 Appendix

2/3

QIS 12J05C01-01E

(2) Setting Password 070.

a. Press the [>] key once. The data display will show “000” with the second digit of 0

flashing.

b. Press the [

š

] key seven times. The data display will show “070” with the second digit of

7 flashing.

c. Press the [ENT] key. The message display will show “HIF.”

d. Press the [YES] key.

(3) Confirming the date and time.

a. Press the [ENT] key. The data display will show the date in day, month, year format.

b. Press the [ENT] key. The data display will show the time in hour, minute, second format.

c. Press the [ENT] key. The message display will show “mA.”

This is the inspection mode. When the message display shows “mA,” check the current outputs.

Each time the [ENT] key is pressed, the value on the data display increase in steps of 4.0 like

“4.0,” and “8.0.” Check the current outputs. Corresponding to the data display, the current

output must be within the range shown in Table 1.

Table 1

Data Display

Current Output (mA DC)

4.0 4

±0.02

12.0 12

±0.02

20.0 20

±0.02

3.3 Temperature

Indication

Check

Following Section 3.2, press the [ENT] key until the message display shows “22kNTC.” in this

state, change the resistance of the decade resistance box and check the data display. The

value on the data display must be within the range shown in Table 2.

Table 2

Temperature (°C)

Decade Box Resistance (:)

Data Display (°C)

-9 99.6k

-9

±0.3

25 22k

25

±0.3

140 653.8

140

±0.3

3.4 Galvanic

Indication

Check

Following Section 3.3, press the [ENT] key until the message display shows “GALVA.” Then

turn SW1 on. Change the current of the DC generator 1 and check the data display. The value

on the data display must be within the range shown in Table 3.

Table 3

Reference (μA)

Data Display (μA)

10 10

±0.06

30 30

±0.1

50 50

±0.2

3.5 Polarograph

Indication

Check

Press the [ENT] key until the message display shows “POLAR.1.” Then turn SW1 off. Change

the current of the DC generator 2 and check the data display. The value on the data display

must be within the range shown in Table 4.