Installation and Wiring 3-4

3-3-2. Wiring other galvanic sensors

Consult the users manual for the color identification

of the sensor cable.

- The temperature sensor has two wires and must

be connected to the terminal 11 and 12.

- The measuring electrode: the cathode must be

connected to terminal 13.

- The reference electrode: the anode must be

connected to terminal 15.

- The overall shield of the cable must be connected

to terminal 14 if there is one available.

- Place the jumper in the GALVANIC position.

3-3-3. Wiring polarographic sensors

Consult the owners manual for the color identifica-

tion of the sensor cable.

- The temperature sensor has two wires and must

be connected to the terminal 11 and 12.

- The measuring electrode: the cathode must be

connected to terminal 17.

- The reference electrode: the anode must be

connected to terminal 18.

- The overall shield of the cable must be connected

to terminal 14 if there is one available.

- Place the jumper in the POLAROGRAPHIC position.

3-3-4. Wiring sensors with Vp connector

Consult the owners manual for the color identifica-

tion of the sensor cable

- The temperature sensor has two wires and must

be connected to terminal 11 and 12

- Polarographic : the measuring electrode

(cathode) must be connected to terminal 17

- Polarographic : the reference electrode (anode)

must be connected to terminal 18

- Galvanic : the measuring electrode (cathode)

must be connected to terminal 13

- Galvanic : the reference electrode (anode) must

be connected to terminal 15

- The overall shield of the cable must be connected

to 14

- Place the jumper in the Polarographic or Galvanic

position

Figure 3-6. Connections DO

Figure 3-7. Terminal indentification labels

A-17 pol. 13 galv.

B-18 pol. 15 galv.

C

D

E-11

F-12

S-14

temp

11

12

12

13

14

14

16

15

13

14

14

16

15

17

11

17

11 Red

12 Blue

15 Core 16 Screen
White Co-axial cable

13 Core 17 Screen
Brown Co-axial Cable

WF10 Cable

14 Overall Screen

TRANSMITTER / CONVERTER

11

12

17

13

15

16

14

A

C

Overall

shield

B

D

Screen

E

Red

White

Blue

Brown

Figure 3-8. Connection of WF10 extension cable and BA10/BP10 junction box

3-3-5. Sensor cable connections using junction

box (BA10) and extension cable (WF10)

Where a convenient installation is not possible using

the standard cables between sensors and trans-

mitter, a junction box and extension cable may be

used. The Yokogawa BA10 junction box and the

WF10 extension cable should be used. These items

are manufactured to a very high standard and are

necessary to ensure that the specifications of the

system are not compromised. The total cable length

should not exceed 30 metres (including DO sensor

cable).