2. commissioning mode, 2-1 output range – Yokogawa DO202 2-Wire Dissolved Oxygen Analyzer User Manual

Page 43

Advertising
background image

IM 12J05C01-01E

5-3 Parameter Setting

5-2-1 Output range

5-2. Commissioning mode

In order to obtain peak performance from the EXA

transmitter, you must set it up for each custom

application. Refer to APPENDIX A-1-2 for stand-

ard configurations and options.

Output range: mA output is set as default

to 0 - 20ppm.

For enhanced resolution in more

stable measuring processes, it may

be desirable to select for example

5-10 ppm range.

Hold

The EXA DO202 transmitter has the

ability to “HOLD” the output during

maintenance periods. This

parameter should be set up to hold

the last measured value, or a fixed

value to suit the process.

Service

This selection provides access to

the service menu.

What follows are pictorial descriptions of typical

frontplate pushbutton sequences for each param-

eter setting function. By following the simple YES/

NO prompts and arrow keys, users can navigate

through the process of setting range, setpoints,

hold, wash and service functions.