4. configuration checklist for do402, 4. configuration checklist for do402 -5 – Yokogawa DO402 Dissolved Oxygen Analyzer User Manual

Page 13

Advertising
background image

IM 12J05D02-01E

Introduction 1-5

Logbook

Software record of important events and diagnostic data. Available through RS485, for use with the

Yokogawa PC402 communication software.

Serial Communication
Bi-directional according to the EIA-485 standard using HART-protocol and PC402 software.

1-4. Configuration checklist for DO402

Standard configuration

Options

Reference for change

Measured variable(s)

Primary inputs

D.O. and Temp

DO range

0- 20 mg/l

any span within 0-50 mg/l

“range”

DO units

mg/l

ppm. % saturation

code 56

Temperature range

0- 50 °C

25 °C minimum span

“range”

Temperature unit

Celsius

Fahrenheit

code 11

Outputs

Analog output

4- 20 mA for DO

0-20 mA or 4-20 mA

code 30

Second output

4- 20 mA for Temp

0-20 mA or 4-20mA

code 30

Output allocation

DO and Temp

DO, Temp, Table, Pl control

code 31

Contact outputs

S1= high at 19.5 mg/l

(4) freely programmable

“setpoint”

S2= low at 1.0 mg/l

code 40. 41, 42, 43

Contact allocation

mg/l and FAIL

mg/l, temp, wash, Pl control, HOLD

code 40- 43

Contact variables

dead time= 0.2 s; hyst= 0.1 mg/l

time: 0- 200 s; hyst 0- 20 mg/l

code 44

Add. contact functions

none

time out alarm

code 47

Control functions

none

Pl on contacts or mA2 output

code 45, 46, 34 33

Digital outputs

none

RS485

code 60

Communication

Digital interface

disabled

RS485

code 60

Communication software disabled

PC402

contact factory

Variables on display

mg/l and temp

%sat, °C, mA1, mA2, SL, ZR, REL

“display”

Burn out

disabled

burn low (3.5)/ high (22) on mA1/ mA2 code 32

Password protection

disabled

for mains/ comm./ serv level

code 52

Autoreturn

return to measure in 10 minutes

enable or disable

code 50

Add. functions in MAINT

disabled

wash start/ setpoint adj

code 51

Diagnostics

Membrane condition

active

enable or disable

code 02

Check on Slope

active

enable or disable

code 02, 22

Check on zero

disabled

enable or disable

code 02, 21, 22

Check on stability

0.05 mg/l per 60 s

0-50 mg/l in 10-600 s

code 20

Compatibility

DO Sensor

DO30

DOX8, DO410, Ingold, DO70G

code 01, 10

Temperature sensor

Pt1000

Pt 100, PB36, 22k NTC

code 10

Membrane thickness

50 µm ( 2 mil)

25 or 50 µm (1 or 2 mil), adj. slope

code 01

Sensor principle

galvanic

galvanic or polarographic *1

code 01, “wiring”

Manual temp. comp.

disabled

disable or enable

code 13

Special Features

Salinity compensation

disabled

comp. for 0- 100 ppt NaCI

code 04

Temperature calibration

none

adjustment +/- 7.5 °C

code 12

Zero calibration

disabled

disable or enable

code 21

Sensor washing

disabled

interval < 36 hours, wash time 0.1-10 min. “Wash”

HOLD during maintenance disabled

hold last or hold fix

“Hold”

Contact during HOLD

disabled

possible on S1,S2, S3 or S4

code 40-42

Atm. press. compensation active

active or manual

Soft fail alarm

disabled

possible for E1..E4, E7..E9, E12, E16, E22 code 53

Logbook

disabled

2 volumes of 50 events

code 61,62

*1: Choose a polarographic sensor for an optical dissolved oxygen sensor.