3. model and suffix codes, 3. model and suffix codes -3 – Yokogawa DO402 Dissolved Oxygen Analyzer User Manual

Page 19

Advertising
background image

IM 12J05D02-01E

Specification 2-3

B. Performance : Temperature (Pt1000, PB36,

22kNTC)

- Linearity

: ± 0.3 ºC

- Repeatability

: ± 0.1 ºC

- Accuracy

: ± 0.3 ºC

Performance : Temperature (Pt100)

- Linearity

: ± 0.4 ºC

- Repeatability

: ± 0.1 ºC

- Accuracy

: ± 0.4 ºC

Note on performance specifications:

The specifications are expressed with sim-

ulated inputs, because the DO402G can be

used with many different sensors with their

unique characteristics.

The following tolerance is added to above

performance.

mA output tolerance : ± 0.02 mA of

"0/4 - 20 mA"

Digital display tolerance : +1 digit

C. Response time

0- 90%

: 10 s

D. Ambient operating temperature

: -10 to +55 °C (14 to 131 ºF)

E. Storage temperature

: -30 to +70 °C (-22 to 158 ºF)

F. Humidity : 10 to 90% RH non-condensing

G. Housing

Case

: Cast aluminium with chemically

resistant coating

Cover

: flexible polycarbonate window.

Case color : off-white (munsell 2.5Y8.4/1.2)

Cover color : moss green. ( munsell

0.6GY3.1/2.0)

Cable entry : via six Pg13.5 nylon glands.

Cable terminals: for up to 2.5 mm

2

finished wires.

Protection

: weather resistant to IP65

standards.

Mounting

: Pipe, wall or panel, using

optional bracket.

H. Data protection

: Non volatile memory for

configuration and logbook, and

lithium battery for clock support.

I. Watchdog timer

: Checks microprocessor

J. Automatic safeguard

: Return to measuring mode

when no keystroke is made for

10 min.

K. Power interruption

: Less than 50 milliseconds no

effect.

L. Operation protection

: 3-digit programmable password.

M. Connection via cables:

The distance between the ensor and trans

mitter can be up to 50 m if the WTB10 junc

tion box is used. This junction box can

not be used for the DO70G, for which the

allowable maximum distance is10 m.

2-3. Model and suffix codes

[Style: S3]

Model

Suffix
code

Option
code

Description

DO402G

----------------- --------- Dissolved Oxygen

Converter

Type

-1

--------- General

Power Supply

Voltage

-1

-2

-5

---------

---------

---------

115V +/-15% AC, 50/60 Hz

230V +/-15% AC, 50/60 Hz

100V +/-15% AC, 50/60 Hz

Language

-J

-E

---------

---------

Japanese

English

Options

Mounting Hardware

Hood

Tag Plate

Conduit Adaptor

Coating

/U

/PM

/H3

/H4

/SCT

/AFTG

/ANSI

/X1

Pipe, wall mounting

bracket (Stainless steel)

Panel mounting bracket

(Stainless steel)

Hood for sun protection

(Carbon steel)

Hood for sun protection

(Stainless steel)

Stainless steel tag plate

G 1/2

1/2 NPT

Epoxy baked finish (*1)

*1: The housing is coated with epoxy resin.