7. wiring other galvanic sensors, 8. wiring the standard optical sensor (do70g), 7. wiring other galvanic sensors -10 – Yokogawa DO402 Dissolved Oxygen Analyzer User Manual

Page 30: 8. wiring the standard optical sensor (do70g) -10

Advertising
background image

IM 12J05D02-01E

3-10 Installation and wiring

3-7. Wiring other galvanic sensors

Consult the users manual for the color identification of the sensor cable and connect temperature com-

pensator, cathode and anode to the terminals: 11, 12, 13 and 15 as described above.

Connect the cable shield to 14 if there is one available. Sensor diagnostics is normally not possible.

Note:

A jumper cable is placed to connect converter terminals 13 and 17.

Make sure that connecting the sensor cable to the IE pin or pin 13 or pulling the sensor cable

does not cause the jumper cable to come loose. Insufficient tightening may cause unstable or

false measurements.

3-8. Wiring the standard optical sensor (DO70G)

Note:

A jumper cable is placed to connect converter terminals 13 and 17.

When a standard optical sensor (DO70G) is to be connected, the cable is not used, so disconnect

it. It is recommended to save the jumper cable for future use—if may be needed when a galvanic

sensor is used.

The color of the wiring of the sensor cable supports the indication of each terminal.

Color of sensor cable

White

Green

Green/Yellow

Clear

Red

Brown

Blue

DO402G

11

12

14

17

18

DOX10

+

F0311.ai

Digital

Communications

12

13

15

16

63 66 65 62 61 95 94 93 92 91

SCREEN

mA2

mA1

SCREEN

TL

TL

17 11

CONT

m

A OUTPUT

14

22

21

18

SCREEN

99

SCREEN 2

23

Sensor Input

Temp.

Wiring optical sensor

mA Output

Contact

Input

Shield

Figure 3-12. Wiring optical sensor