3-6. user interface, 3-6. user interface -26 – Yokogawa DO402 Dissolved Oxygen Analyzer User Manual

Page 62

Advertising
background image

IM 12J05D02-01E

5-26 Parameter setting

5-3-6. User interface

Code 50

*RET

When Auto return is enabled, the converter reverts to the measuring mode

from anywhere in the configuration menus, when no button is pressed during

the set time interval of 10 minutes.

Code 51

*MODE

The adjustment of the contact setpoints, and the manual operation of the

wash system can be setup for operation in the maintenance mode. (Through

the closed front cover). In maintenance mode the pressure can be entered

(M. PRESS). If manual pressure is enabled, a warning is given when starting

manual-, water and air calibration.

Code 52

*PASS

Passcodes can be set on any or all of the access levels, to restrict access to

the instrument configuration.

Code 53

*Err.04

Error message configuration. Two different types of failure mode can be set.

Hard fail gives a steady FAIL flag in the display, and a continuous contact

closure. All the other contacts (controls) are inhibited (except HOLD contacts),

and a Fail signal is transmitted on the outputs when enabled in code 32.

Soft fail gives a flashing FAIL flag in the display, and the relay contacts are

pulsed. The other contacts (controls) are still functional, and the controller

continues to work normally. The call for maintenance is a good example of

where a SOFT fail is useful. A warning that the regular maintenance is due,

should not be used to shut down the whole measurement. An extra choice is

added for Errors set to Soft Fail. It can be set to LCD + Fail contact (current

situation) or to LCD only.

Code 54

*E12

Overrange alarm, set limit for supersaturation permissible before FAIL alarm is

triggered.

Code 55

*CALL.M

Call for maintenance is a contact trigger to signal that the system has been in

service for longer than the set time without calibration. The user can set up to

250 days as a routine service interval.

Code 56

*UNIT

Select mg/l, ppm or % saturation as the main unit, to suit the application.

Advertising