3. setting passcodes, 3. setting passcodes -3 – Yokogawa DO402 Dissolved Oxygen Analyzer User Manual

Page 35

Advertising
background image

IM 12J05D02-01E

Operation 4-3

4-3. Setting passcodes

In Service Code 52, EXA users can set passcode protection for each one of the three operating levels,

or for any one or two of the three levels. This procedure should be completed after the initial commis-

sioning (setup) of the instrument. The passcodes should then be recorded safely for future reference.

When passcodes have been set, the following additional steps are introduced to the configuration and

programming operations:

Maintenance

Press MODE key. The display shows 000 and *PASS*

Enter a 3-digit passcode as set in Service Code 52 to obtain access to the Maintenance Mode

Commissioning
Press

*

key. The display shows 000 and *PASS*

Enter a 3-digit passcode as set in Service Code 52 to obtain access to the Commissioning Mode.

Service

From the commissioning menu, select *SERV by pressing YES key. The display shows 000 and *PASS*

Enter a 3-digit passcode as set in Service Code 52 to obtain access to the Service Mode.

NOTE:

See Service Code 52 for the setting of passcodes.