3-1. parameter specific functions, 3-1. parameter specific functions -14 – Yokogawa DO402 Dissolved Oxygen Analyzer User Manual

Page 50

Advertising
background image

IM 12J05D02-01E

5-14 Parameter setting

5-3. Notes for guidance in the use of service coded settings

5-3-1. Parameter specific functions

Code 01

*S.TYPE

Select 0 for DO30 sensor or other galvanic sensor. Select 1 for SM31, other

polarographic sensor and optical dissolved oxygen sensor.

*I.CELL

After selecting sensortype you need to select the nominal cell current. With a

galvanic sensor 3 choices are offered.

Select 0 for DO30 thick (50 µm) membrane with a default of 3.75 µA.

Select 1 for DO30 thin (25 µm) membrane with a default of 7.5 µA.

Select 9 for user definable. This can be set to suit any galvanic sensor.

With polarographic sensor 2 choices are offered.

Select 0 for SM31 sensor with a nominal cell current of 50 nA.

Select 9 for other polarographic sensor, with the nominal cell current to be user

defined.

Select 9 for the DO70G sensor and input the nominal cell current of 60 nA.

General Note: Dissolved oxygen sensors fall into two categories – galvanic and

polarographic. Yokogawa offer both types:

Settings for Yokogawa Sensors

DO30

DO30

SM31.

DO70G

Membrane Thick

Thin Sterilizable Fluorescence

*S.TYPE

0

0

1

1

* I.CELL

0

1

0

9

Code 02

* CHECK This code is used to set up appropriate diagnostics. In most cases a zero

calibration is not used, the default check, therefore, set to “off”. When zero

check enabled, the maximum permissible zero offset should be set as a signal

current.

When slope check enabled, the maximum as minimum calibrated slope values

should be set as % of the nominal value.

For the galvanic sensors, a membrane check can be enabled. This is an

impedance test to ensure that the membrane is intact.

The membrane check is not available for optical and polarographic sensors, so

the membrane check cannot be enabled or disabled.

Code 04

*SAL.TY

Set 0 (salinity compensation disabled) when measuring water not including

salinity components. Set 1 (salinity compensation enabled) when manually

calibrated to salinity compensation. This is intended for accurately displaying

saturated % concentration when measuring sample water. Set 1 (salinity

compensation enabled) and then enter NaCl concentration using *[CL-]. For

this concentration, enter salinity concentrations in actual sample water used to

figure out salinity compensation values.

NOTE:

The instrument will not perform salinity compensation simply by setting the

salinity compensation to “1” (enabled) in Code 04, but requires the NaCl

concentration. For measurement in consideration of the salinity of a sample,

first, manually calibrate with values corrected for salinity and enter the NaCl

concentration in Code 04. (See Section 6.4, “Calibration method using manual

calibration method”)

Advertising