2. commissioning mode, 2. commissioning mode -5 – Yokogawa DO402 Dissolved Oxygen Analyzer User Manual

Page 41

Advertising
background image

IM 12J05D02-01E

Parameter setting 5-5

5-2. Commissioning mode

In order to obtain peak performance from the EXA converter, you must set it up for each custom applica-

tion. Refer to section 1-4 for standard configurations and options.

*SETP

Alarms are set by default

S1 - high process alarm

S2 - low process alarm

S3 - WASH

S4 - FAIL

The setpoints are at arbitrary default value. Therefore, you must set these to

meaningful values, or set them to off. (See service codes 40 to 49 and user

interface codes 50 to 59).

*RANGE

mA output 1 is set as default to 0 - 20 mg/l

mA output 2 is set as default to 0 - 50 °C

For enhanced resolution in more stable measuring processes, it may be desir-

able to select 0 - 5 mg/l range, for example, and maybe 0 - 25 °C temperature

range.

Service codes 30 to 39 can be used to choose other output parameters on mA

output 2.

*HOLD

The EXA converter has the ability to “hold” the output during maintenance peri-

ods. This parameter should be set up to hold the last measured value, or a fixed

value to suit the process.

*WASH

The EXA can be set up to control a wash cleaner. When using this function, the

timings must be configured for interval, wash and recovery periods.

*SERV

This selection provides access to the service menu.

What follows are pictorial descriptions of typical frontplate pushbutton sequences for each parameter set-

ting function. By following the simple YES/NO prompts and arrow keys, users can navigate through the

process of setting range, setpoints, hold, wash and service functions.