3-2. temperature functions, 3-3. calibration functions, 3-2. temperature functions -16 – Yokogawa DO402 Dissolved Oxygen Analyzer User Manual

Page 52: 3-3. calibration functions -16

Advertising
background image

IM 12J05D02-01E

5-16 Parameter setting

5-3-2. Temperature functions

Code 10

*T.SENS

Select temperature sensor to suit the measuring probe. Refer to instructions

with DO sensor for which temperature sensor is used.

Code 11

*T.UNIT

Select °C or °F for temperature display.

Code 12

*T.ADJ

Adjust offset in temperature measurement when the sensor is at a stable

known value.

Code 13

*T.MAN

Enable and set manual temperature compensation.

After the manual temperature set menu has been displayed, enter

temperature signals and press the [ENT] key to set. If the manual temperature

compensation is enabled, the temperature set menu is displayed in

maintenance mode.

The maximum and minimum slope values that may be entered depend on the nominal current from serv-

ice code 01.

Example

: When nominal current = 3.75 µA the maximum and minimum slope calibration values

that may be entered are:

Maximum

3.75 x 2 = 7.5 µA

Minimum

3.75 x 0.4 = 1.5 µA

5-3-3. Calibration functions

Code 20

*∆T.SEC

Set stability parameters for calibration. A greater time, and/or a smaller step

gives

*∆mg/l

greater stability before a calibration can be accepted. Use caution, however, to

avoid very long calibration times.

Code 21

*0.CAL

In the DO30G sensor or other galvanic sensors and polarographic sensors,

the zero point fluctuates less than the span. Therefore, with the zero check, a

zero calibration is not required for these sensors provided that they are used in

environments with high-concentration dissolved oxygen such as aerobic tanks.

The zero calibration is recommended when installing sensors, replacing

membranes, or performing measurement in anaerobic tanks with low-

concentration dissolved oxygen.

The two-point calibration (zero and span) is necessary for the DO70G optical

dissolved oxygen sensor because of its operation principle. Select 1 to enable

zero calibration.

NOTE:

The zero point is reset to the default when 0 is set (zero calibration disabled)

even after calibration.

Code 22

*ZERO

Zero current values can be entered manually. If a zero calibration is made in

maintenance mode, a zero current value will be automatically reset.

NOTE:

A zero current value is displayed only when a zero calibration is enabled in

code 21.

*SLOPE

Slope values can be manually entered with current values. If a span calibration

is made in maintenance mode, slope values will be automatically reset.

NOTE:

Slope values can be displayed with current values. When span check functions

are enabled in code 02, the above current values are converted to % values

and compared with the set values.

Slope % = current sensitive current /reference sensitive current

Advertising