Yokogawa DO402 Dissolved Oxygen Analyzer User Manual

Page 91

Advertising
background image

IM 12J05D02-01E

Appendix 2 11-3

3/4

QIS 12J05D02-01E

Table 3

Temperature (°C)

Resistance ()

Data Display (°C)

5

1019.5

5 ±0.3

25

1097.3

25 ±0.3

45

1174.7

45 ±0.3

3.6 Input Current Indication Check

Following Section 3.5, press the [ENT] key until the message display shows “GALVA1.”

Turn SW1 on. Change the current of the standard current generator and check the data

display. The value on the data display must be within the range shown in Table 4.

Table 4

Reference (µA)

Data Display (µA)

0

0 ±0.05

8

8 ±0.05

16

16 ±0.05

Press the [ENT] key until the message display shows “GALVA2.” Change the current of

the standard current generator and check the data display. The value on the data display

must be within the range shown in Table 5.

Table 5

Reference (µA)

Data Display (µA)

10

10 ±0.1

30

30 ±0.2

50

50 ±0.2

Turn SW1 off. Press the [ENT] key until the message display shows “POLAR1.” Change

the current of the standard current generator and check the data display. The value on

the data display must be within the range shown in Table 6.

Table 6

Reference (nA)

Data Display (nA)

0

0 ±0.2

80

80 ±0.2

160

160 ±0.2

Press the [ENT] key until the message display shows “POLAR2.” Change the current of

the standard current generator and check the data display. The value on the data display

must be within the range shown in Table 7.

Table 7

Reference (nA)

Data Display (nA)

100

100 ±0.2

300

300 ±1

500

500 ±1