3-7. communication setup, 3-8. general, 3-9. test and setup mode – Yokogawa DO402 Dissolved Oxygen Analyzer User Manual

Page 64: 3-7. communication setup -28, 3-8. general -28, 3-9. test and setup mode -28

Advertising
background image

IM 12J05D02-01E

5-28 Parameter setting

5-3-7. Communication setup

Code 60

*COMM.

The settings should be adjusted to suit the communicating device connected to

the RS485 port.

*SET.

*ADDR.

For the Yokogawa PC402 software package, the default settings match the

software as shipped.

Code 61

*HOUR

The clock/calendar for the logbook is set for current date and time as

reference.

*MINUT

*SECND

*YEAR

*MONTH

*DAY

Code 62

*ERASE

Erase logbook function to clear the recorded data for a fresh start. This may

be desirable when re-commissioning an instrument that has been out of service

for a while.

5-3-8. General

Code 70

*LOAD

The load defaults code allows the instrument to be returned to the default set

up with a single operation. This can be useful when wanting to change from

one application to another.

5-3-9. Test and setup mode

Code 80

*TEST

Not used

Note:

Attempting to change data in service code 80 and above without the proper instructions and

equipment, can result in corruption of the instrument setup, and will impair the performance of the

unit.