Troubleshooting 8-1

8. TROUBLESHOOTING

This chapter describes the countermeasures for failures, classifying the cases into three categories:

dissolvedoxygen converter failure, detection of failure with the self-diagnosis function, and abnormal

measured values. The causes for abnormal measured values are not limited to equipment failures. if an

abnormal phenomenon occurs, first check the following items:

• Is the property of the measuring solution different from normal ?

• Is the dissolvedoxygen sensor properly installed?

8-1. Measures in the case of converter operation failure

8-1-1. No dissolved-oxygen converter operation

If the converter does not operate even when the power is supplied, one cause may be a blown fuse

inside the converter.

Examine the fuse by turning off the power supply (see Subsection 3-3-1 on page 3-5).

If the fuse is blown, replace it with a new one. If fuses frequently blow and the cause is not clear, request

an inspection by Yokogawa.

If the fuse is normal, examine the wiring system.

8-1-2. Operation key or display failure

If the operation keys do not operate smoothly or the display fails (e.g., a missing character segment),

repair of the printed circuit board (digital board) (replacement with a new one) is required.

After the printed circuit board is replaced, operation checks and parameter settings are necessary.

Contact Yokogawa and request board-replacement work.