5. system configuration, 5-1. dissolved oxygen sensor, 5. system configuration -6 – Yokogawa DO402 Dissolved Oxygen Analyzer User Manual

Page 14: 5-1. dissolved oxygen sensor -6

Advertising
background image

IM 12J05D02-01E

1-6 Introduction

1-5. System Configuration

(For DO30G)

Output Signal:

4 to 20 m A DC or

0 to 20 m A DC

DO30G

DOX8A

DOX8 W

DO70G

DO402G

WTB10-DO3

Alarm Contact Output
4-Contact Point,High/Low Alarm

Cleaning Time,Alarm Contact

HOLD Contat

100V AC±15%
115V AC±15%
230V AC±15%

Guide Pipe

 PH8HG

Submersion Type Holder

 DOX8HS

Suspension Type Holder

 HH350G

F01.EPS

Floating Ball Holder

 PB350G

Dissolved Oxygen Sensor

Spere Parts Set

Holders, Holders with Cleaning System

Dissolved Oxygen Converter

PB360G

Calibration Set

DOX10

(For DO30G)

(For DO30G)

(For DO30G)

(For DO70G)

Select pin terminal

for DO402G.

No Cleaning Jet Cleaning

Terminal Box

Power Supply Unit

1-5-1. Dissolved Oxygen Sensor

The DO402G can be used with a variety of sensors. Some specific programming may be necessary to

match the converter with the sensor.

a. DO70G sensor

This sensor features temperature compensation with a 22k NTC RTD. The nominal output of this

sensor is 60 nA in air under reference conditions. Before use, enter the service mode. Select service

code 01 (*S.TYPE) and set it to 1 (polarographic sensor), then press the ENT key. Select *1.CELL2.9

and set the nominal output to 60 nA. Next, select service code 10 (*T.SENS) and set it to 3 (22k

NTC). Then select service code 21 (*0.CAL) and set it to 1 (zero calibration enabled) because two-

point calibration by air-saturated water calibration and zero calibration are required.

For details, refer to GS12J05D04-01E.

b. Model DOX8SM sensor

This sensor is a galvanic sensor with a PB36 type temperature compensator and a 25 micron

membrane (1 mil).

This sensor has a nominal output current of 7.5µA in ambient air under reference conditions. In

Service code 01 selection 1 must be programmed when this sensor is used. The membrane check

is inactive with this sensor unless external solution ground is used. ( metal holder or sample line )

c. Model DO30G sensor

This sensor is essentially the same as the DOX8SM sensor, but it has a 50 micron membrane,

fixed sensor cable, Pt1000 RTD temperature compensation and integral solution ground that allows

membrane checking. The nominal output of this sensor is 3.75 µA in air under reference conditions.

In Service item ”*I.CELL” selection 0 must be programmed when this sensor is used.

For details, refer to GS12J05D03-01E.

d. Model DO30 sensor

This sensor is specified in GS12J6K4 and it can be supplied with 50 or with 25 micron membrane.

Normally 50 micron membrane is used and no reprogramming is necessary. Otherwise Service

code 01 setting must be changed from 0 to 1.