Yokogawa DO402 Dissolved Oxygen Analyzer User Manual
Page 59
Advertising

IM 12J05D02-01E
Parameter setting 5-23
Code Display
Function
Function detail
X
Y
Z
Default values
Contacts Settings
44
*D.TIME
Dead Time
Delay after setpoint passed 0- 2.0 s
0.2 s
*P.HYST
Hysteresis
Process value hysteresis
0-50 mg/l or 0-300 %
0.1 mg/l
*T.HYST
Temp hyst
Temperature hysteresis
0- 5 °C (0- 12 °F)
0.5 °C
45
*RANGE
Prop. control
Proportional range 0- 50 mg/l
5.0 mg/l
*PER.
Duty cycle
Period/ Duty cycle : 5- 100 s
10.0 s
*FREQ.
Frequency
Frequency control 50 - 120 per min.
70 p.m.
46
*tI.CNT
Integral time
Int. time for contacts 10 - 1999 s
100 s
47
*EXPIR
Time out alarm
Disable time out alarm
0
0
Enable time out alarm
1
*tE. mn
time out alarm after 0.5 - 199.9 min
15 min
48-49
Not used
Advertising