3-2. access to terminal and cable entry, 3-2. access to terminal and cable entry -6, High voltage compartment – Yokogawa DO402 Dissolved Oxygen Analyzer User Manual

Page 26

Advertising
background image

IM 12J05D02-01E

3-6 Installation and wiring

3-3-2. Access to terminal and cable entry

Terminals 1 and 2 on the bottom terminal strip are used for the power supply. Guide the power cables
through the gland closest to the power supply terminals. The terminals will accept wires of 2.5 mm

2

(14

AWG). Use cable finishings if possible.

Connect the wires as indicated in the wiring diagram (refer to figure 3-6).

High Voltage Section

A

B

C

D

E

F

D

E

F

CONT

ACT (S3, S4/FAIL)

OUTPUT CABLE (S)

CONT

ACT (S1, S2)

OUTPUT CABLE (S)

POWER CABLE

A

B

C

POWER CABLE

for DO70G or

COMMUNICATION

INPUT CONTACT

SENSOR CABLE (S)

ANALOG OUTPUTS

CABLE

Suitable for cables with an outside diameter between 6 - 12 mm (0.24 - 0.47 in.)

Figure 3-7. Glands to be used for cabling

71

S4

S3

S2

S1

C NC NO

72 73 51 52 53 41

43 31

33

42

32

FUSE

100
115
230

250VAC; T

3

1

2

C NC NO C NC NO C NC NO

VAC
VAC
VAC

200 mA
200 mA
100 mA

Relay Contacts

Power Supply

12

13

15

16

63 66 65 62 61 95 94 93 92 91

SCREEN

mA2

mA1

SCREEN

TL

TL

17 11

Sensor Inputs

mA Outputs

Digital

Communications

Contact

Input

CONT SENSOR mA OUTPUT RS485

14

22

21

- - - optical sensor and polarographic sensor galvanic sensor

High voltage compartment

G N L

18

Liquid

earth

Shield

Temp

SCREEN

99

SCREEN 2

23

250VAC

5A

100VA

250VDC

5A

50W

Figure 3-8. Input and output connections