Yokogawa DO402 Dissolved Oxygen Analyzer User Manual

Page 55

Advertising
background image

IM 12J05D02-01E

Parameter setting 5-19

Code Display

Function

Function detail

X

Y

Z

Default values

mA Output Functions

30

*mA

mA output

Output 1 is 0- 20 mA

0

1.1

Output 1 is 4- 20 mA

1

Output 2 is 0- 20 mA

0

Output 2 is 4-20 mA

1

31

*OUTP.F

Output function

Process value on mA1

0

0.2

Table on mA1

1

Process value on mA2

0

Table on mA2

1

Temperature on mA2

2

PI control on mA2

3

*D/R

Control action

Reverse acting PI-control

0

0

Direct acting PI control

1

32

*BURN

Burn-out

No burn-out function on mA1

0

0.0

Burn low on fail on mA1

1

Burn high on fail on mA1

2

No burn out function on mA2

0

Burn low on fail on mA2

1

Burn high on fail on mA2

2

33

*RGmA2

Prop. range

Proportional band: 0- 50 mg/l

5.0 mg/l

0-300 % saturation

34

*tImA2

Integral time

Integral time in seconds 10 - 1999 s

100 s

35

*TABL1

Table for mA1

21 point table: Process value at

0, 5, .. 100 % of span

none

36

*TABL2

Table for mA2

21 point table: 0, 5, .. 100 %

none

37

*DAMP

Output damping

damping 0- 120 s

0 s

38

*SWTCH

Range 1B

Disabled

0

0

Enabled

1

39

Not used

Note

To enable wash function, set Service Code 38 to 0, Service Code 42 to 4.0.0, and Service Code

51 to 0.1.0.