Appendix, 1. setpoint, 2. range – Yokogawa DO402 Dissolved Oxygen Analyzer User Manual

Page 85: 3. hold, 4. wash, Appendix -1, 1. setpoint -1, 2. range -1, 3. hold -1, 4. wash -1

Advertising
background image

IM 12J05D02-01E

Appendix 10-1

10. APPENDIX

Adjustable limits and defaults for settings in Commissioning Level

10-1. Setpoint

10-2. Range

Note:

The range has a minimum span of: 1.0 mg/l, 1.0 ppm or 10 %sat.

In case of a temperature output 25 ºC or 50 ºF

10-3. Hold

10-4. Wash

Variable

Default

Lower limit

Upper limit

*OX1

19.5 mg/l

0.0 mg/l

50.0 mg/l

0.0 ppm

50.0 ppm

0.0 %sat

300 %sat

*OX2

1.0 mg/l

0.0 mg/l

50.0 mg/l

0.0 ppm

50.0 ppm

0.0 %sat

300 %sat

*OX3

10.0 mg/l

0.0 mg/l

50.0 mg/l

0.0 ppm

50.0 ppm

0.0 %sat

300 %sat

*OX4

10.0 mg/l

0.0 mg/l

50.0 mg/l

0.0 ppm

50.0 ppm

0.0 %sat

300 %sat

*Sp.mA2

10.0 mg/l

0.0 mg/l

50.0 mg/l

0.0 ppm

50.0 ppm

0.0 %sat

300 %sat

* T1

25.0 ºC/77 ºF

0.0 ºC/32 ºF

50 ºC/122 ºF

* T2

25.0 ºC/77 ºF

0.0 ºC/32 ºF

50 ºC/122 ºF

* T3

25.0 ºC/77 ºF

0.0 ºC/32 ºF

50 ºC/122 ºF

* T4

25.0 ºC/77 ºF

0.0 ºC/32 ºF

50 ºC/122 ºF

Variable

Default

Lower limit

Upper limit

Range 1:

0% = 0.00

0.0 mg/l

50.0 mg/l

Process

100% = 19.99

0.0 ppm

50.0 ppm

0.0 %sat

300 %sat

Range 2:

0% = 0.00

0.0 mg/l

50.0 mg/l

Process

100% = 19.99

0.0 ppm

50.0 ppm

0.0 %sat

300 %sat

Temperature

0% = 0 ºC / 32 ºF

0.0 ºC / 32 ºF

50 ºC / 122 ºF

100% = 50 ºC / 122 ºF

Variable

Default

Lower limit

Upper limit

mA fix 1

10.00 mA

0 or 4 mA

20 mA

mA fix 2

10.00 mA

0 or 4 mA

20 mA

Variable

Default

Lower limit

Upper limit

Interval

6 hours

0.1 hours

36.0 hours

Wash time

0.5 minutes

0.1 minutes

10.0 minutes

Recovery time

0.5 minutes

0.1 minutes

10.0 minutes