Yokogawa DO402 Dissolved Oxygen Analyzer User Manual
Page 65

IM 12J05D02-01E
Parameter setting 5-29
Code Display
Function
Function detail
X
Y
Z
Default values
General
70
*LOAD
Load defaults
Reset configuration to default values
71-79
Not used
Code Display
Function
Function detail
X
Y
Z
Default values
Communication
60
*COMM.
Communication
Set communication
Off
0
0.1
Off
Set communication
On
1
Communication write enable
0
write
Communication write protect
1
protect
*SET.
Baud rate & parity
Baud rate
1200
0
3.1
2400 1
4800 2
9600
3
9600
Parity
Off
0
Odd
1
Odd
Even 2
*ADDR.
Network address
Set address 00 to 15
00
61
*HOUR
Clock setup
Adjust to current date and time using
*MINUT
>, ^ and ENT keys
*SECND
*YEAR
*MONTH
*DAY
62
*ERASE
Erase logbook
Press YES to clear logbook data
63-69
Not used
Code Display
Function
Function detail
X
Y
Z
Default values
Test and setup mode
80
*TEST
Test and setup
Not used