7-2 maintenance

7-2. Periodic maintenance for the EXA DO402G converter

The DO402G converter requires very little periodic maintenance. The housing is sealed to IP65 (NEMA

4X) standards, and remains closed in normal operation. Users are required only to make sure the front

window is kept clean in order to permit a clear view of the display and allow proper operation of the

pushbuttons. If the window becomes soiled, clean it using a soft damp cloth or soft tissue. To deal with

more stubborn stains, a neutral detergent may be used.

NOTE:

Never used harsh chemicals or solvents. In the event that the window becomes heavily stained or

scratched, refer to the parts list (Chapter 9) for replacement part numbers.

When you must open the front cover and/or glands, make sure that the seals are clean and correctly fit-

ted when the unit is reassembled in order to maintain the housing’s weatherproof integrity against water

and water vapor. The DO measurement uses small signal currents and may otherwise be prone to prob-

lems caused by exposure of the circuitry to condensation.

7-3. Fuse Replacement

This section describes how to replace the power supply fuse.

• Use only a fuse of the specified current, voltage and type ratings to prevent fire.

• Power to the instrument must be turned off at the external breaker before replacing the fuse. After

replacing the fuse, install a high-voltage shield plate before turning the power on.

• Do not short circuit the fuse holder.

Specified Ratings of the Power Supply Fuse

The power supply fuse (recommended fuse) used in the DO402G is as follows.

• 100/115 V AC version

Maximum rated voltage: 250 V

Maximum rated current: 200 mA

Compliance:

UL, CSA, VDE, Japan’s Electrical Appliance and material Safety Law

Part number:

A1105EF

• 230 V AC version

Maximum rated voltage: 250 V

Maximum rated current: 100 mA

Compliance:

UL, CSA, VDE, Japan’s Electrical Appliance and material Safety Law

Part number:

A1103EF