IM 12J05D02-01E

Installation and wiring 3-11

DO70G Wiring Diagram Example

DO402G

DOX10

Detail-1

Detail-2

DO70G

Detail-1(DO402G)

Detail-2(DOX10)

Connections of the sensor cable and DO402G terminals

are shown in the following table.

Connections of the sensor cable and DOX10 terminals

are shown in the following table.

DO70G (Wiring color)

Red

Clear

Green/Yellow

Green

White

DO402G (Terminal No.)

18
17
14
12
11

DO70G (Wiring color)
BROWN
BLUE

DOX10 (terminal No.)
+ BROWN
− BLUE

Fugire 3-13. Wiring optical sensor (DO70G)

3-9. Wiring polarographic sensors

Consult the owners manual for the color identification of the sensor cable and connect temperature com-

pensator, cathode and anode to terminals 11, 12, 17 and 18 in this sequence.

Connect the cable shield (when provided) to terminal 14.

Note:

A jumper cable is placed to connect converter terminals 13 and 17.

When a polarographic sensor is to be connected, the cable is not used, so disconnect it.

It is recommended to save the jumper cable for future use—if may be needed when a galvanic

sensor is used.

F0311.ai

Digital

Communications

12

13

15

16

63 66 65 62 61 95 94 93 92 91

SCREEN

mA2

mA1

SCREEN

TL

TL

17 11

CONT

m

A OUTPUT

14

22

21

18

SCREEN

99

SCREEN 2

23

Sensor Input

Temp.

Wiring polarographic sensor

mA Output

Contact

Input

Shield

Figure 3-14. Wiring polarographic sensor

3-10. Wiring RS485 signal

It is necessary to use screening/shielding on the output signal cables. Screw (M3) 99 is used to connect

the shielding.